Powerheart® G5 Automated External Defibrillator
70-00569-02 G
2-7
7: Prepare the AED for the next rescue
7: Prepare the AED for the next rescue
After transferring the patient to the emergency medical personnel, close the lid of the AED. Prepare
the AED for the next rescue:
1.
Open the lid.
2.
Optional: Retrieve the rescue data stored in
the internal memory of the AED. See the
AED Manager User’s Guide
for details.
3.
Connect a new adult pads package to the AED.
See the
Defibrillation Pads Instructions for Use
for details.
4.
Verify that the pad connection indicator is off.
If the indicator is on, make sure that the pad
connector is properly attached to the AED.
5.
Verify that there is adequate charge (1)
remaining in the battery. If the battery charge
is low (2), replace the battery.
6.
Verify that the service indicator is off.
7.
Close the lid.
8.
Verify that the Rescue Ready indicator
is green.
(1)
(2)
R
E A D Y
R E
S CUE
Summary of Contents for Powerheart G5