70-01047-01 B
Product Safety and Setup
30
Maintenance
Once per shift: clean
Clean the ECG device at least once per shift, or as needed.
!
Caution: Equipment damage.
Do not use ether, bleach, acetone, benzene, or similar solvents to
clean the ECG device, cables, or electrodes. Use only these
cleaning agents: 3M™ 23H, PDI® Nice Pak® Sani-System, or
Virex II™.
!
Caution: Equipment damage.
Do not immerse the ECG device, cables, or electrodes in any type
of fluid.
If the ECG device is immersed in, or comes in contact with, large
quantities of fluid, immediately discontinue use and contact a
qualified service technician.
!
Caution: Equipment damage.
Do not hot sterilize ECG device, cables, or electrodes.
Do not
sterilize the patient cable.
Exterior
To clean the exterior of the ECG device:
◆
Use a clean damp cloth to apply a disinfectant listed above.
Patient cable and reusable electrodes
See the
Accessory List
for Cardiac Science-approved patient cables
and electrodes. To clean and disinfect:
◆
Use a clean damp cloth to apply a disinfectant listed above.