CAPINTEC, INC.
CAPTUS
®
3000
AUTO CALIBRATION
Calibration is done with Cs137 and Eu152 standard sources.
1. Click the
button or press
Alt+A
on Figure 5-1 Quality Assurance Main
Screen. Figure 5-3 Auto Calibration Screen will appear. When the procedure is started,
the spectrum appears in the upper half of the screen as it is being collected and displays
the measurement time elapsed in the upper right corner. Note that the
button at the top of the screen is enabled only after calibration has been done at least
once on that day.
Figure 5-3 Auto Calibration Screen
2. Note that in the
Select Detector
frame,
PROBE
is selected by default. To change the
detector to Well, click the radio button marked
WELL
.
3. Click the
button or press
Alt+S
to start the calibration procedure. The
screen appears as shown in Figure 5-4.
Figure 5-4
July 14
QUALITY ASSURANCE
5-5
Summary of Contents for CAPTUS 3000
Page 2: ......
Page 14: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 22: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 86: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 124: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 266: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 284: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 292: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 304: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 334: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 336: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...
Page 337: ......
Page 338: ...CAPINTEC INC CAPTUS 3000 This page intentionally left blank...