The following table shows different behaviours with different warm-up values:
Warm-Up value
Effetto
0
The device doesn
’
t generate any voltage but, in any case, carries out the measurements.
>=1
<=32
The device generates a voltage for a period in seconds equal to the value inserted +450ms
(
approximately
).
>32
The device sets-up the warm-up time to 32s and generates a voltage for a period of
approximately 32.45s.
Decimal value
The device round down the inserted value.
<0
The device sets-up the warm-up time to 32s and generates a voltage for a period of
approximately 32.45s.
•
Save the new configuration clicking on
“
Write
”
button.
7.
Device pre-set and use mode.
a.
Wireless Mode:
No setup operation is needed. Typically, the system is configured in factory, so the device is already associated to the system gateway. The
device is in STANDBY mode (
refer to
) for which is necessary to start it with the TEST command (
refer to
Otherwise, in case the device is in FACTORY RESET mode (
refer to
), that means it's ready for connecting to an existing
system, in order to associate it, make reference to the
WineCap System - User Manual R30
software manual. It is necessary to use the
software on the PC connected to the gateway that will be linked with the device.
b.
USB Logger Mode:
For this operation mode, stand-alone with data downloads through USB, the connection with the PC and the
is necessary to modify the device 's operation mode. The sampling interval must be set with the device in STAND-ALONE
(
refer to
alone USB datalogger installation.
mode and automatically, the device 's clock is aligned with the PC's clock, in order to assure the temporal
reference of the sample.
Sampling operations start may be selected disconnecting the USB cable or giving the proper command with the magnetic key (
refer to
Stand-alone USB datalogger installation.
). More details on device 's connection/disconnection through the USB cable are available on the
manual.
c.
On field transition from USB to Wireless datalogger:
This transition is practicable in field, during the sampling period, using the wireless network association command. When the association is
done, the datalogger becomes a wireless datalogger and, besides sending new measures to the gateway, starts a download process towards
the same gateway of the measures acquired during the stand-alone period (
refer to
Picture 9 - Stand-alone device - Status table