background image

This In-Service training session is of vital importance to you and your staff. It is the 

only method of ensuring staff competency in the operation and understanding of this 

equipment. The level of understanding gained during this In-Service can have a direct 

effect on patient and staff safety. 

Please afford the time and attention required to ensure that you and your staff fully  

understand the training provided.

Daily Maintenance

1. Check and refill detergent reservoir if used.
2. Clean detergent reservoir if using diluted solution.
3. Refill alcohol reservoir if necessary.
4. Wipe basins with lint-free cloth.
5. Drain condensation from air tank (pull ring on tank).

Weekly Maintenance

1. Lubricate the connector O-ring in the leak test  
    hookup with silicone oil.

Three Months

1. Replace air filter.

Six Months

1. Replace 0.2 micron water filter every 6 months or  
    if below 40 psi and run waterline disinfection.
2. Replace 1 micron and 0.45 micron water filter  

every 6 months or if below 40 psi.

3. Replace vapor management filter (if applicable).

MAINTENANCE

TROUBLESHOOTING

Error Reports

1. Review interruptions caused by sensors: Lid, fluid,  
    air flow, basin level, reservoir levels.
2. Error Messages - See troubleshooting guide in  
    DSD User Manual.

Run the Reprocessing Cycle

1. Place floating lid on the basin. Verify endoscope or hookup does not contact floating lid and close reprocessor lid.
2. Test HLD with test strips to confirm potency.
3. Press “STATION SELECT” and choose Station A or B.
4. Press “ID DATA” button and enter ID data: serial number, operator ID, patient ID, physician ID.
5. Select disinfection program and press START.
6. DSD Prompts: LCG Test Pass? “PRESS ENTER,” if test strip passed. Press “CANCEL” if failed; dump and replace HLD.
7. Verify fluid flows through hookups, endoscope channels and from distal end and that there are no leaks.

Complete the Reprocessing Cycle

1. Open reprocessor lid when LCD Screen displays “Completed”. Confirm endoscope ID number on print out is correct, 
    and “Completed” is documented.
2. Remove floating lid, disconnect hookup and remove endoscope. If any hookup component connection is loose or 
    disconnected during the cycle, endoscope must be reprocessed.
3. Store hookup and reprocessed endoscope with all removable parts detached.

End of Day Shutdown

1. Close incoming water line.
2. Sanitize upper basins and basin lids with an EPA-registered sanitizer.

SYSTEM OPERATION

Endoscope Reprocessing System

DSD-201

© 2019 Medivators Inc.  All rights reserved.  50097-064 Rev F

Medivators Customer and Technical Support Toll Free:  Phone: 1-800-444-4729 

·

 www.medivators.com

This Medivators guide, form, checklist or procedure is designed for use solely as a customer education tool, and is not intended to replace or in any way modify the Medivators User and Service 

Manuals. While Medivators training may be used in support of a facility’s overall competency program, it shall not constitute certification of facility personnel. Medivators shall in no event,  

be held responsible for a facility’s proper performance of endoscope reprocessing protocols or a facility staying current and compliant with ongoing instructional changes or corresponding 

training updates. Facility owners and operators of Medivators equipment are fully responsible for complying with industry standards and manufacturer’s proper use instructions.  

INTERCEPT™ and RAPICIDE™ are trademarks of Medivators Inc.

All other company or product names referenced in this document are trademarks of their respective owner.

FOR FACILITY USE ONLY

Staff Name:

Training Date:

Trainer:

IN-SERVICE GUIDE

DSD-201 & DSD201LT

Reviews: