Quick Guide - V5 / V8 OXYGEN CONCENTRATOR USER GUIDE AND MAINTENANCE – Page 1 

 

V5 / V8 OXYGEN CONCENTRATOR FRONT VIEW 

 

Control Panel 
 

 

 

 

ALARMS 

 

Power off alarm sound: 

Concentrator connected with power but the unit is in a 
state of power off with alarm sound.  
Check the power cord for good connection or that the 
external power supply has been turned on. 

 

Oxygen purity alarm sound: 

Alarm will sound in addition to red indicator light if 
oxygen concentration 

purity 

%

 drops.

 

 

Alarm Coding: 

10000 

– overheat warning 

02000 

– Valve Defective Warning 

00300 

– over pressure warning (blockage) 

00040 

– low pressure warning (leakage) 

 

CONTROL PANEL 
 

Display screen 

  

Reset button                                        Oxygen flow valve 

 

Indicator lights 

 

Stop clock/Timer buttons 

 

INDICATOR LIGHTS 

 

Oxygen purity 

Green

 –

 Optimal

 

 

Red

 –

 Oxygen 

purity % drops

 

Trouble indicator 
Off

 - Working good 

Red

 - Defective 

Power indicator 

Green

 – Power on 

Red

 – Power off 

alarm 

 

 

 

DO NOT USE STOP CLOCK/TIMER BUTTONS 

 

OXYGEN FLOW CONTROL VALVE 

 
 
 

    

 

 

 

 

OXYGEN OUTLET 

 

 

         

 

 

 
 

 

 
 
 
 
 
 

 
 

 

On/off Switch 

Oxygen outlet 

 

 

 

Humidifier bottle Concavity 

V5: 0-5 l/min  
V8: 0-8 l/min  

Setting flow rate: 

Turn the flow control valve: 
Counter-clockwise: on 
Clockwise: off 

 
 

Oxygen tubing 

Connect nasal 
cannula/Mask direct to 
oxygen outlet or via the 
humidifier bottle if in 
use. 

 

 

 

MESU 

DO NOT OVER TIGHTEN FLOW VALVE KNOB WHEN CLOSING! 
DO NOT TAKE THE FLOW VALVE BALL TOO HIGH ABOVE THE 
MAXIMUM NUMBER SHOWN! 