8. Calibration
– 98 –
2
Select the sensor unit.
If multiple sensor units can be used, first select a sensor unit by touching that sensor unit
button.
3
Remove the grid unit.
If a sensor unit with removable grid unit is selected in
step 2
, remove the grid unit.
If
[START]
is touched while the grid unit is attached, it will display an error, and
calibration will not be performed.
4
Set times (number of exposures).
Set the number using or .
Or, touch the frame and enter the value from the ten key window.
5
Select the bucky time.
Touch to select from the list.
This setting can only be made when using a movable grid.
Calibration will not be performed properly unless the X-ray irradiation time is within the
bucky time.
6
Set the exposure parameters.
Set the number using or .
Or, touch the frame and enter the value from the ten key window or parameter table.
7
Set the timeout for warning.
Set the period from the completion of calibration until a warning is displayed to perform
the next calibration. Touch
[hour]/[day]/[week]
to select the unit of time.
Touch or to set the display duration.
Or, touch the frame and enter the value from the ten key window.
A warning will not be displayed if the time is set to
0
.
8
Enter a memo as required.
Touch the frame to display the software keyboard and enter the memo in 64 or fewer
alphanumeric characters.
Warning on calibration
The warning on calibration will be displayed as a yellow
[ERR]
mark blinking on the
lower right side of the message area.
Although exposures can be performed even when the warning is displayed, you must
remember to calibrate the instrument later, as the warning will disappear when another
error occurs.
Summary of Contents for CXDI-1
Page 9: ...Starting Up Shutting Down 1 Starting Up Shutting Down the CXDI ...
Page 15: ...Exposure and Output 2 Taking an Image 3 Output Control Process Viewer ...
Page 38: ...2 Taking an Image 30 ...
Page 60: ...5 Viewing Multiple Images Multi View Screen 52 ...
Page 70: ...6 Browsing Completed Studies Study List 62 ...
Page 71: ...Processing the Image 7 Processing the Image QA Mode Screen ...
Page 103: ...CXDI Adjustment 8 Calibration 9 Self test ...
Page 110: ...8 Calibration 102 ...
Page 116: ...9 Self test 108 ...
Page 117: ...Troubleshooting 10 Before Calling a Service Engineer 11 Questions and Answers ...
Page 136: ...11 Questions and Answers 128 ...
Page 156: ...B Using the Rejection Reason Input Function 148 ...
Page 162: ...C Using the Sensor Unit Attach Detach Function 154 ...
Page 169: ......