New GentleYAG
Cryogen Spray Alignment Procedure
Candela Corporation
8503-01-0845, Revision A
Candela Corporation Proprietary
Page 7 of 10
Distance gauge spot size
“Typical”
spray duration
to minimally fill the inside
ring spot.
3 mm
30 ms
8 mm
30 ms
10 mm
30 ms
12 mm
40 ms
15 mm
50 ms
18 mm
60 ms
Results: If the spot doesn’t fill or a leak is noted, the handpiece assembly should be
replaced or contact Candela Technical Support.