the sum of the excitation strength and the flag value. The following table summarizes diagnostic
codes that indicate warning flags:
Table 9-5: Diagnostic code ranges
Diagnostic code
range
1
Excitation
only
Low-
amplitude
warning
flag present
2
High-
amplitude
warning
flag present
2
Low-
frequency
warning
flag present
2
High-
frequency
warning
flag present
2
0 – 255
ü
256 – 511
ü
512 – 767
ü
768 – 1023
1024 – 1279
ü
1280 – 1535
ü
ü
1536 – 1791
ü
ü
1792 – 2047
2048 – 2303
ü
2304 – 2559
ü
ü
2560 – 2815
ü
ü
1
Diagnostic codes outside these ranges may occur when hardware, firmware, or CRBasic programming errors
occur.
2
Amplitude warning flags and frequency warning flags have the following restrictions:
l
A measurement can have a low-amplitude warning flag or high-amplitude warning flag, but not both.
l
A measurement can have a low-frequency warning flag or a high-frequency warning flag, but not both.
l
A measurement can have both an amplitude warning flag and a frequency warning flag.
9.7.3 DVW Toolbox errors
As shown in
(p. 38), DVW Toolbox Error Indicators, DVW Toolbox indicates
measurement errors in the following ways:
l
Information icon
in the Channel. Click icon for more information.
l
Red or flashing-red measurement values
VWIRE 305