![Campbell SkyVUE 8 Product Manual Download Page 55](http://html1.mh-extra.com/html/campbell/skyvue-8/skyvue-8_product-manual_3808856055.webp)
Table 5-3: Summary of message ID and descriptions
Message_ID_x type
Description
000
No message
001
Campbell Scientific Message 1, no sky condition, no profile data
002
Campbell Scientific Message 2, no sky condition, profile data, 1600 range
bins, 5 m resolution
003
Campbell Scientific Message 3, sky condition, no profile data
004 (default)
Campbell Scientific Message 4, sky condition and profile data, 1600 range
bins, 5 m resolution
101
CL31 Message 1, 770 range bins, 10 m resolution
102
CL31 Message 1, 385 range bins, 20 m resolution
103
CL31 Message 1, 1500 range bins, 5 m resolution
104
CL31 Message 1, 770 range bins, 5 m resolution
105
CL31 Message 1, No profile data
106
CL31 Message 1, Full SkyVUE 8 output, 1600 range bins, 5 m resolution
107
CL31 Message 2, 770 range bins, 10 m resolution
108
CL31 Message 2, 385 range bins, 20 m resolution
109
CL31 Message 2, 1500 range bins, 5 m resolution
110
CL31 Message 2, 770 range bins, 5 m resolution
111
CL31 Message 2, No profile data
112
CL31 Message 2, Full SkyVUE 8 output, 1600 range bins, 5 m resolution
113
CT25K Message 1
114
CT25K Message 6
5.1.6 Measurement and message intervals
The message interval is the time, in seconds, between the automatic message transmissions. It
can be set between 2 to 600 s (
0
gives polled messages). The default is 10, meaning a message
will be sent automatically every 10 seconds. The message interval must be a multiple of the
measurement period.
SkyVUE™8 (CS136) LIDAR Ceilometer
43
Summary of Contents for SkyVUE 8
Page 1: ...Revision 08 2020 Copyright 2019 2020 Campbell Scientific CSL I D 1301...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 108: ......
Page 112: ...SkyVUE 8 CS136 LIDAR Ceilometer 99...
Page 115: ......