
SI-111 and SI-111SS Precision Infrared Radiometers
8
7.4.1 Sensor Body Temperature Measurements
7.4.1.1 Therm109() Instruction
CRBasic dataloggers measure the sensor body temperature (°C) using the
Therm109()
instruction:
Therm109
(
Dest, Reps, SEChan, Mult, Offset
)
Use the default values for
Mult
and
Offset,
which output the temperature in °C.
7.4.2 Thermopile Detector
7.4.2.1 VoltDiff() Instruction
CRBasic dataloggers measure the thermopile detector using the
VoltDiff()
instruction:
VoltDiff
Dest, Reps, Range, DiffChan, RevDiff, SettlingTime, Integ/Fnotch,
The
Range
value should be
mV200
(CR6, CR1000X),
mV2_5
(CR800 series,
CR1000),
mV34
(CR300 series),
mV20
(CR3000, CR5000), or
mV50
(CR9000(X)). The
Integ/Fnotch
parameter is often set to the 60 Hz option.
Use the default values for the
RevDiff, SettlingTime, Mult
, and
Offset
parameters.
7.4.3 Target Temperature Calculations
These calculations are entered in CRBasic as expressions. Calculate the target
temperature using the following steps:
1.
Calculate slope
2.
Calculate intercept
3.
Convert sensor body temperature (°C) to Kelvin.
4.
Calculate target temperature (K)
5.
If desired, convert target temperature (K) to degrees Celsius
Use Equation 1 to calculate slope:
m = mC2 • SBTempC^2 + mC1 • S mC0
(1)
Where,
m = slope
SBTempC = sensor body temperature in °C
mC2 = slope polynomial coefficient (C2) from the calibration sheet
mC1 = slope polynomial coefficient (C1) from the calibration sheet
mC0 = slope polynomial coefficient (C0) from the calibration sheet
Use Equation 2 to calculate intercept:
b = bC2 • SBTempC^2 + bC1 • S bC0
(2)