PVS5120-Series Portable Samplers
30
8.2.2 Time-Based Control
Time-based control allows the PVS5120 to run as a stand-alone sampler.
Status values are not reported unless a datalogger is used for monitoring the
sampler. A datalogger can be used to monitor the sampler over RS-485 and
uses either the PakBus® or Modbus® protocol. An MD485 is also required
Monitoring Sampler Status via RS-485
(p. C-1)
).
If running with time-based control, the user must sync the
clock of the sampler. This is described above in Section
(p. 10)
.
1. Once connected to the VSC100, click the
Sampler Settings
tab (FIGURE
).
a. Set the
Control Type
to
Time
.
b. If using PakBus® to monitor the sampler over RS-485, set the
PakBus Address
.
c. If using Modbus® to monitor the sampler over RS-485, set the
Modbus Address
.
d. Set the
Number of Sampler Bottles
.
e. Set the
Samples Per Bottle
. It can range from
1
to
999
. This setting
is ignored if
Never Full
is set to
On/Enable
.
f. Typically leave the
Never Full Operation
disabled.
g. If the Auto Deploy routine has been run, the
Purge Time
should
already be set. If not using Auto Deploy, enter the amount of time the
sampler hose is evacuated before and after the sample is collected.
h. If a
Water Present Sensor
is present, click to install. This sensor is
wired to the sampler control cable. See Section
(p. 24)
, for more information.
A
CAUTION
Summary of Contents for PVS5120 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 70: ...Appendix A Sample Transport Velocity A 2...
Page 78: ...Appendix B Example Programs B 8...
Page 86: ...Appendix C Monitoring Sampler Status via RS 485 C 8...
Page 96: ...Appendix G Generic Modbus Control G 4...
Page 100: ...Appendix H ISO5667 Conformity H 4...
Page 101: ......