CSAT3B Three-Dimensional Sonic Anemometer
25
TABLE 7-1. CSAT3B Settings and Status Values in
Device Configuration Utility
Subscreen
Setting or Status
Value
Options
Description
Internal
Temperature
-
Temperature inside CSAT3B block
Internal Relative
Humidity
-
Relative humidity inside CSAT3B block.
Change desiccant if greater than 50% (see
Inclinometer
Pitch
-
Pitch angle of CSAT3B head (see Section
(p. 27)
)
Inclinometer Roll
-
Roll angle of CSAT3B head (see Section
(p. 27)
)
Communication
Settings
SDM Address
1 through 14
3
1/
Unique address for SDM device
CPI Address
1 through 120
30
1/
Unique address for CPI device
Communication
Port Protocol
Disable Both
CPI Enabled
1/
RS-485 Enabled
Identifies whether the RS-485/CPI port will be
disabled, enabled for CPI, or enabled for
RS-485
CPI Baud Rate
Auto
1/
1000 kbps
500 kbps
250 kbps
125 kbps
50 kbps
Baud rate for CPI communications
(in most circumstances this setting should
remain set to
Auto
)
RS-485 Baud
Rate
1200
2400
4800
9600
19200
38400
57600
115200
1/
230400
Baud rate for RS-485 communications
Unprompted
Output Port
Disabled
1/
USB Port
RS-485 Port
Identifies the port to output unprompted data
Unprompted
Output Rate
10 Hz
1/
20 Hz
50 Hz
100 Hz
Identifies the rate at which to output
unprompted data