![Campbell CRVW3 Instruction Manual Download Page 40](http://html1.mh-extra.com/html/campbell/crvw3/crvw3_instruction-manual_489101040.webp)
CRVW3 3-Channel Vibrating-Wire Datalogger
27
button to save these settings onto the computer as an xml file (*.xml). A
filename is suggested which includes the device type and the date. You can
change the name of the file to something different if you wish. Later on, you
can use the
Read File
button to retrieve the device settings that were saved to
an xml file, and re-apply them to the device with the
Apply
button.
When applying settings to a device using the
Apply
button, the
Save
screen
appears, allowing you to review and save the settings. Use the
Save
button to
save the *.xml file at that time, if desired.
8.1.6 Testing the Measurement Operation of the CRVW3
You can use the
Device Configuration Utility
in the lab to confirm the
measurement operations of the CRVW3. Once you have configured the
CRVW3 as desired, the CRVW3 will restart itself and automatically begin
measuring its sensors at the interval you have specified. The data is stored
immediately as each measurement is made.
When you first connect to the CRVW3, the following tabs are available:
Settings Editor
,
Logger Control
,
Data Monitor
,
Send OS
,
Troubleshoot
,
and
Terminal
. If any changes are made to settings within the
Settings Editor
tab and its sub-tabs, all of these tabs except
Settings Editor
will disappear
until those changes are either applied or canceled.
Use the
LoggerControl
tab to check the real-time clock of the device. The
CRVW3’s clock can be synchronized with the PC clock using the
Set Clock
button. If the
LoggerControl
tab is unable to obtain the time from the
CRVW3, there are connectivity problems that need to be resolved (see Section
Troubleshooting CRVW3 Networks
Use the
Data Monitor
tab to examine the data being stored in the final storage
tables and utility tables of the device. Click on a table’s name in the left pane,
and its latest stored record will be displayed in the right pane. The
Status
table
shows general information about the CRVW3 and its operation. The
Public
table shows information about specific variables in use by the current program
configuration. The
VW_Data
table is the main storage table. This is where a
permanent history of sensor readings is stored. The
DataTableInfo
table gives
details about final storage tables in use by the device. If calibration settings are
Summary of Contents for CRVW3
Page 2: ......
Page 3: ......
Page 7: ......
Page 13: ...Table of Contents vi ...
Page 121: ...Appendix A Updating CRVW3 Firmware A 4 ...
Page 131: ...Appendix E Engineering Output and Calibration E 4 ...
Page 133: ......
Page 134: ......