A-1
Appendix A. Updating CRVW3
Firmware
From time to time, Campbell Scientific will release updates to the CRVW3
firmware (also known as the
Operating System
or
OS
) which will improve the
operating stability of the product and may introduce minor feature
enhancements. Once published, these are available for download from the
Campbell Scientific website:
The version of the OS currently active in the CRVW3 is shown in the
Status
table (as
OSVersion
) which can be viewed in the
Data Monitor
tab of
DevConfig
or in the
Table Monitor
in the
Connect Screen
of
LoggerNet
. The
version of the OS is also shown in the
Main
tab on the
Settings Editor
tab of
DevConfig
. A date (
OSDate
) and signature (
OSSignature
) are also shown to
help ensure proper installation. Compare your current OS version to that
available on the website to determine whether an update is required. Campbell
support personnel may also provide you with a CRVW3 OS under certain
circumstances.
The OS is packaged as an *.obj file. This file can be sent to the CRVW3 using
DevConfig
or
LoggerNet
. When the CRVW3 receives an *.obj file, the update
of the firmware is triggered automatically.
Any attempt to send a new OS to the CRVW3 will reset the final
storage on the device. You should collect any data needed from
the CRVW3 before making any attempt to update its OS. The
settings of the device will remain intact.
To install the OS with
DevConfig
:
1.
Open
DevConfig
and choose the
CRVW Series
for the device.
2.
Choose the
Send OS
tab.