Section 7. Installation
154
8. Set
Mode
=
4
to start second part of calibration.
a.
Mode
=
5
(automatic) during second point calibration.
b.
Mode
=
6
(automatic) when calibration is complete.
7.8.1.5 FieldCal() Demonstration Programs
FieldCal()
has the following calibration options:
•
Zero
•
Offset
•
Two-point slope and offset
•
Two-point slope only
•
Zero basis (multi-put zero)
These demonstration programs are provided as an aid in becoming familiar with
the
FieldCal()
features at a test bench without actual sensors. For the purpose of
the demonstration, sensor signals are simulated by the CR800 excitation channel.
To reset tests, use the support software
File Control
(p. 431)
menu commands to
delete .cal files, and then send the demonstration program again to the CR800.
Term equivalents are as follows:
"offset"
=
"y
‐
intercept"
=
"zero"
"multiplier"
=
"slope"
=
"gain"
7.8.1.5.1 Zero or Tare (Option 0)
Zero option simply adjusts a sensor's output to zero. It does not affect the
multiplier.
Case: A sensor measures the relative humidity (RH) of air. Multiplier is known to
be stable, but sensor offset drifts and requires regular zeroing in a desiccated
chamber. The following procedure zeros the RH sensor to obtain the calibration
report shown. Use the external keyboard / display or software numeric monitor to
change variable values as directed.
Table 22.
Calibration Report for Air RH Sensor
Parameter
Argument at Deployment
Argument at 30-Day
Service
mV
output
1000
mV
1050
mV
KnownRH
(desiccated
chamber)
0
%
0
%
Multiplier
0.05
% / mV
0.05
% / mV
Offset
-50
%
-52.5
%
RH
reading
0
%
0
%
Summary of Contents for CR850
Page 2: ......
Page 4: ......
Page 6: ......
Page 26: ...Table of Contents 26...
Page 30: ...Section 2 Cautionary Statements 30...
Page 32: ...Section 3 Initial Inspection 32...
Page 35: ...Section 4 Quickstart Tutorial 35 Figure 2 Wiring panel...
Page 55: ...Section 4 Quickstart Tutorial 55 Figure 24 PC200W View data utility...
Page 78: ...Section 5 System Overview 78...
Page 80: ...Section 6 CR800 Specifications 80...
Page 267: ...Section 7 Installation 267 Figure 84 Running average signal attenuation...
Page 268: ...Section 7 Installation 268...
Page 384: ...Section 8 Operation 384 Figure 113 Using the keyboard display...
Page 387: ...Section 8 Operation 387 Figure 116 Real time custom...
Page 388: ...Section 8 Operation 388 8 8 1 3 Final Storage Tables Figure 117 Final storage tables...
Page 389: ...Section 8 Operation 389 8 8 2 Run Stop Program Figure 118 Run Stop Program...
Page 390: ...Section 8 Operation 390 8 8 3 File Display Figure 119 File display...
Page 396: ...Section 8 Operation 396...
Page 402: ...Section 9 Maintenance 402...
Page 450: ...Section 11 Glossary 450...
Page 504: ...Appendix A CRBasic Programming Instructions 504...
Page 526: ...Appendix B Status Table and Settings 526...
Page 530: ...Appendix C Serial Port Pinouts 530...
Page 536: ...Appendix E FP2 Data Format 536...
Page 550: ...Appendix F Other Campbell Scientific Products 550...
Page 565: ......