Section 8. Operation
357
FIGURE 82: Input voltage rise and transient decay
CRBasic Measurement Settling Times
SettlingTime
Argument
Integ
Argument
Resultant
Settling Time
1
0
250
450 µs
0
_50Hz
3 ms
0
_60Hz
3 ms
integer ≥
100
integer
μs entered in
SettlingTime
argument
1
450 µs is the minimum settling time required to meet CR800 resolution
specifications.
Settling Errors
When sensors require long lead lengths, use the following general practices to
minimize settling errors:
•
Do not use wire with PVC-insulated conductors. PVC has a high
dielectric constant, which extends input settling time.
•
Where possible, run excitation leads and signal leads in separate shields
to minimize transients.
•
When measurement speed is not a prime consideration, additional time
can be used to ensure ample settling time. The settling time required
can be measured with the CR800.
•
In difficult cases, settling error can be measured as described in
Measuring Settling Time
(p. 320).
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......