FIGURE 61: Strain Gage Shunt
Calibration Finish
7.7.11.6.4 FieldCalStrain() Quarter-Bridge Zero
Continuing from
FieldCalStrain() Quarter-Bridge Shunt Example
(p. 231),
keep the
249 kΩ resistor in place to simulate a strain. Using the CR1000KD
Keyboard/Display or software numeric monitor, change the value in variable
Zero_Mode
to
1
to start the zero calibration as shown in figure
Zero Procedure
Start
(p. 232).
When
Zero_Mode
increments to
6
, zero calibration is complete as
shown in figure
Zero Procedure Finish
(p. 232).
FIGURE 62: Zero Procedure Start
FIGURE 63: Zero Procedure Finish