Section 7. Installation
99
Note
Lightning strikes may damage or destroy the CR800 and
associated sensors and power supplies.
In addition to protections discussed in
,
use of a simple lightning rod and low-
resistance path to earth ground is adequate protection in many installations. .
FIGURE 31: Lightning Protection Scheme
7.3.2 Single-Ended Measurement Reference
Low-level, single-ended voltage measurements (<200 mV) are sensitive to ground
potential fluctuation due to changing return currents from
12V
,
SW12
,
5V
, and
C1
–
C4
terminals. The CR800 grounding scheme is designed to minimize these