Section 8. Operation
372
8.1.3.1 Pulse Measurement Terminals
P Terminals
•
Input voltage range =
–20 to 20 V
If pulse input voltages exceed
±
20 V, third-party external-signal conditioners
should be employed. Under no circumstances should voltages greater than 50 V
be measured.
C Terminals
•
Input voltage range =
–8 to 16 Vdc
C
terminals configured for pulse input have a small 25 ns input RC-filter time
constant between the terminal block and the CMOS input buffer, which allows for
high-frequency pulse measurements up to 250 kHz and edge counting up to 400
kHz. The CMOS input buffer recognizes inputs ≥3.
8 V as being high and inputs
≤1.2 V as being low.
Open-collector (bipolar transistors) or open-drain (MOSFET) sensors are
typically measured as frequency sensors.
C
terminals can be conditioned for open
collector or open drain with an external pull-up resistor as shown in figure
Connecting Switch Closures to C Terminals Configured for Control. The pull-up
resistor counteracts an internal 100 kΩ pull
-down resistor, allowing inputs to be
pulled to >3.8 V for reliable measurements.
8.1.3.2 Low-Level Ac Measurements — Details
Related Topics:
•
Low-Level Ac Input Modules — Overview
(p. 395)
•
Low-Level Ac Measurements — Details
(p. 372)
•
Pulse Input Modules — List
(p. 562)
Low-level ac (sine-wave) signals can be measured on
P
terminals. Sensors that
commonly output low-level ac include:
•
Ac generator anemometers
Measurements include the following:
•
Counts
•
Frequency (Hz)
•
Running average
Rotating magnetic-pickup sensors commonly generate ac voltage ranging from
thousandths of volts at low-rotational speeds to several volts at high-rotational
speeds. Terminals configured for low-level ac input have in-line signal
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......