Section 8. Operation
346
measurements. The first measurement determines the range to use. It is made
with a 250 µs integration on the ±5000 mV range. The second measurement is
made using the range determined from the first. Both measurements use the
settling time entered in the
SettlingTime
parameter. Auto-ranging optimizes
resolution but takes longer than a measurement on a fixed range because of the
two-measurement sequences.
An auto-ranged measurement will return
NAN
("not a number") if the voltage
exceeds the range picked by the first measurement. To avoid problems with a
signal on the edge of a range,
AutoRange
selects the next larger range when the
signal exceeds 90% of a range.
Use auto-ranging for a signal that occasionally exceeds a particular range, for
example, a type-J thermocouple measuring a temperature usually less than 476
°
C
(
±
25 mV range) but occasionally as high as 500
°
C (
±
250 mV range).
AutoRange
should not be used for rapidly fluctuating signals, particularly signals
traversing multiple voltage ranges rapidly. The possibility exists that the signal
can change ranges between the internal range check and the actual measurement.
Analog Voltage Input Ranges and Options
Range Code
Description
mV5000
measures voltages between ±5000
mV
mV2500
1
measures voltages between ±2500
mV
mV250
2
measures voltages between ±250 mV
mV25
2
measures voltages between ±25 mV
mV7_5
2
measures voltages between ±7.5 mV
mV2_5
2
measures voltages between ±2.5 mV
AutoRange
3
datalogger determines the most
suitable range
1
Append with
C
to enable common-mode null / open-input detect and set
excitation to full-scale (~2700 mV) (Example:
mV2500
)
2
Append with
C
to enable common-mode null / open-input detect (Example:
mV25C
)
3
Append with
C
to enable common-mode null / open-
input detect on ranges ≤
±250 mV, or just common-mode null on ranges > ±250 mV (Example:
AutoRangeC
)
Input Limits / Common-Mode Range
Related Topics:
• Voltage Measurements — Specifications
•
Voltage Measurements — Overview
(p. 65)
•
Voltage Measurements — Details
(p. 345)
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......