Section 8. Operation
318
Ac Noise Rejection on Large Signals
1
Ac-Power Line
Frequency
Measurement
Integration
Time
CRBasic
Integration
Code
Default
Settling
Time
Maximum
Recommended
Settling Time
2
60 Hz
250 μs • 2
_60Hz
3000 μs
8330 μs
50 Hz
250 μs • 2
_50Hz
3000 μs
10000 μs
1
Applies to analog input voltage ranges
mV2500
and
mV5000
.
2
Excitation time and settling time are equal in measurements requiring excitation. The CR800 cannot excite
VX
excitation terminals during A-to-D conversion. The one-half-cycle technique with excitation limits the length of
recommended excitation and settling time for the first measurement to one-half-cycle. The CR800 does not
prevent or warn against setting a settling time beyond the one-half-cycle limit. For example, a settling time of up
to 50000 µs can be programmed, but the CR800 will execute the measurement as follows:
1. CR800 turns excitation on, waits 50000 µs, and then makes the first measurement.
2. During A-to-
D, CR800 turns off excitation for ≈170 µs.
3. Excitation is switched on again for one-half cycle, then the second measurement is made.
Restated, when the CR800 is programmed to use the half-cycle 50 Hz or 60 Hz rejection techniques, a sensor does
not see a continuous excitation of the length entered as the settling time before the second measurement — if the
settling time entered is greater than one-half cycle. This causes a truncated second excitation. Depending on the
sensor used, a truncated second excitation may cause measurement errors.
Signal Settling Time
Settling time allows an analog voltage signal to settle closer to the true magnitude
prior to measurement. To minimize measurement error, signal settling is needed
when a signal has been affected by one or more of the following:
•
A small transient originating from the internal multiplexing that connects
a CR800 terminal with measurement circuitry
•
A relatively large transient induced by an adjacent excitation conductor
on the signal conductor, if present,because of capacitive coupling during
a bridge measurement
•
Dielectric absorption. 50 Hz or 60 Hz integrations require a relatively
long reset time of the internal integration capacitor before the next
measurement.
The rate at which the signal settles is determined by the input settling time
constant, which is a function of both the source resistance and fixed-input
capacitance (3.3 nfd) of the CR800.
Rise and decay waveforms are exponential. Figure
Input Voltage Rise and
Transient Decay
(p. 319)
shows rising and decaying waveforms settling closer to the
true signal magnitude, V
so
. The
SettlingTime
parameter of an analog
measurement instruction allows tailoring of measurement instruction settling
times with 100 µs resolution up to 50000 µs.
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......