Section 7. Installation
230
FIGURE 59: Quarter-Bridge Strain
Gage with RC Resistor Shunt
FieldCalStrain() Calibration
'This program example demonstrates the use of the FieldCalStrain() instruction by measuring
'quarter-bridge strain-gage measurements.
Public
Raw_mVperV
Public
MicroStrain
'Variables that are arguments in the Zero Function
Public
Zero_Mode
Public
Zero_mVperV
'Variables that are arguments in the Shunt Function
Public
Shunt_Mode
Public
KnownRes
Public
GF_Adj
Public
GF_Raw
'----------------------------- Tables ----------------------------
DataTable
(CalHist,NewFieldCal,50)
SampleFieldCal
EndTable
'//////////////////////////// PROGRAM ////////////////////////////
BeginProg
'Set Gage Factors
GF_Raw = 2.1
GF_Adj = GF_Raw
'The adj Gage factors are used in the calculation of uStrain
'If a calibration has been done, the following will load the zero or
'Adjusted GF from the Calibration file
LoadFieldCal
(True)
Summary of Contents for CR800 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 34: ......
Page 54: ......
Page 92: ......
Page 310: ......
Page 446: ...Section 8 Operation 446 8 11 2 Data Display FIGURE 100 CR1000KD Displaying Data ...
Page 448: ...Section 8 Operation 448 FIGURE 102 CR1000KD Real Time Custom ...
Page 449: ...Section 8 Operation 449 8 11 2 3 Final Storage Data FIGURE 103 CR1000KD Final Storage Data ...
Page 450: ...Section 8 Operation 450 8 11 3 Run Stop Program FIGURE 104 CR1000KD Run Stop Program ...
Page 452: ...Section 8 Operation 452 FIGURE 106 CR1000KD File Edit ...
Page 456: ......
Page 462: ......
Page 523: ...Section 11 Glossary 523 FIGURE 116 Relationships of Accuracy Precision and Resolution ...
Page 524: ......
Page 526: ......
Page 556: ......
Page 558: ......
Page 560: ......
Page 597: ......