Section 8. Operation
455
FIGURE 105: Wide Sweep, High Noise
Minimize Resonant Decay
A narrow-swept range ensures minimal decay of the resonant response prior to
measurement. Gage response starts to decay as soon as the frequency sweep
moves past the resonant frequency. Observing this decay is difficult because the
swept frequency overwhelms the resonant-frequency response while excitation is
still active. Because wider excitation sweeps take longer, the resonant response
decays for a longer time before the CR6 can measure it. The resulting resonant
amplitude is smaller. The figure
Wide Sweep, High Noise
(p. 453)
shows this effect,
with the resonant frequency response decreased by about a factor of two when
compared with the narrow sweep.
Prevent Spectral Leakage
Matching the swept excitation to the expected resonant frequency range prevents
spectral leakage
(p. 602)
from complicating the spectral analysis. Vibrating wire
sensors are usually optimized for a single resonant frequency to overwhelm
harmonic and sub-harmonic responses, so spectral leakage usually has little
Summary of Contents for CR6 Series
Page 2: ......
Page 4: ......
Page 6: ......
Page 32: ......
Page 36: ......
Page 38: ......
Page 76: ...Section 5 Overview 76 FIGURE 20 Half Bridge Wiring Example Wind Vane Potentiometer ...
Page 80: ...Section 5 Overview 80 FIGURE 23 Pulse Input Wiring Example Anemometer ...
Page 136: ......
Page 454: ...Section 8 Operation 454 FIGURE 104 Narrow Sweep High Noise ...
Page 459: ...Section 8 Operation 459 FIGURE 106 Vibrating Wire Sensor Calibration Report ...
Page 535: ...Section 8 Operation 535 8 11 2 Data Display FIGURE 121 CR1000KD Displaying Data ...
Page 537: ...Section 8 Operation 537 FIGURE 123 CR1000KD Real Time Custom ...
Page 538: ...Section 8 Operation 538 8 11 2 3 Final Storage Data FIGURE 124 CR1000KD Final Storage Data ...
Page 539: ...Section 8 Operation 539 8 11 3 Run Stop Program FIGURE 125 CR1000KD Run Stop Program ...
Page 541: ...Section 8 Operation 541 FIGURE 127 CR1000KD File Edit ...
Page 546: ......
Page 552: ......
Page 610: ...Section 11 Glossary 610 FIGURE 137 Relationships of Accuracy Precision and Resolution ...
Page 612: ......
Page 648: ......
Page 650: ......
Page 688: ......
Page 689: ......