SECTION 13. CR510 MEASUREMENTS
13-16
13.5 RESISTANCE MEASUREMENTS
REQUIRING AC EXCITATION
Some resistive sensors require AC excitation.
These include the 207 Relative Humidity Probe,
soil moisture blocks, water conductivity sensors,
and wetness sensing grids. The use of DC
excitation with these sensors can result in
polarization, which will cause an erroneous
measurement, and may shift the calibration of
the sensor and/or lead to its rapid decay.
The AC half bridge Instruction 5 (incorporated
into the 247 conductivity probe) reverses
excitation polarity to provide ion depolarization
and, in order to minimize the time excitation is
on, grounds the excitation as soon as the signal
is integrated (Figure 13.5-1). The slow
integration time should never be used with a
sensor requiring AC excitation because it
results in the excitation lasting about 1.5 times
as long, allowing polarization to affect the
measurement.
INFLUENCE OF GROUND LOOP ON
MEASUREMENTS
When measuring soil moisture blocks or water
conductivity, the potential exists for a ground
loop which can adversely affect the
measurement. This ground loop arises
because the soil and water provide an alternate
path for the excitation to return to CR510
ground, and can be represented by the model
diagrammed in Figure 13.5-2.
FIGURE 13.5-1. AC Excitation and Measurement Sequence for AC Half Bridge
FIGURE 13.5-2. Model of Resistive Sensor with Ground Loop
Summary of Contents for CR510
Page 2: ...This is a blank page ...
Page 4: ...This is a blank page ...
Page 10: ...This is a blank page ...
Page 44: ...SECTION 1 FUNCTIONAL MODES 1 14 This is a blank page ...
Page 56: ...SECTION 3 INSTRUCTION SET BASICS 3 8 ...
Page 104: ...SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES 8 8 This is a blank page ...
Page 128: ...SECTION 10 PROCESSING INSTRUCTIONS 10 8 This is a blank page ...
Page 144: ...SECTION 12 PROGRAM CONTROL INSTRUCTIONS 12 10 This is a blank page ...
Page 172: ...SECTION 14 INSTALLATION AND MAINTENANCE 14 10 This is a blank page ...
Page 176: ...APPENDIX A GLOSSARY A 4 This is a blank page ...
Page 184: ...This is a blank page ...
Page 188: ...APPENDIX D DATALOGGER INITIATED COMMUNICATIONS D 4 This is a blank page ...
Page 196: ...APPENDIX F MODBUS ON THE CR10 X AND CR510 F 4 This is a blank page ...
Page 197: ...APPENDIX G TD OPERATING SYSTEM ADDENDUM FOR CR510 CR10X AND CR23X MANUALS ...
Page 198: ...This is a blank page ...
Page 200: ...This is a blank page ...
Page 206: ...TABLE DATA ADDENDUM AD 6 This is a blank page ...
Page 222: ...TD ADDENDUM SECTION 1 FUNCTIONAL MODES AD 1 8 This is a blank page ...
Page 238: ...TD ADDENDUM SECTION 8 PROCESSING AND PROGRAM CONTROL EXAMPLES AD 8 10 This is a blank page ...
Page 240: ...This is a blank page ...