Section 8. Processing and Math Instructions
8-27
4
Full bridge strain gage, 2 gages parallel to
+ε
, the other 2 parallel to
−ε
:
µε =
−
10
6
V
GF
r
5
Full bridge strain gage, half the bridge has 2 gages parallel to
+ε
and
−ε
: the other half
+νε
and
−νε
:
µε
ν
=
− ⋅
+
2 10
1
6
V
GF
r
(
)
Code
Configuration
6
Full bridge strain gage, one half
+ε
and
−νε
, the other half
−νε
and
+ε
.:
µε
ν
ν
=
− ⋅
+ −
−
2 10
1
1
6
V
GF
V
r
r
[(
)
(
)]
Source
The source variable array for the measurement(s), the input is expected as
millivolts out per volt in (the result of the full bridge instruction with a
multiplier of 1 and an offset of 0.
BrZero
The variable array that holds the unstrained reading(s) in millivolts out per
volt in.
GF
Gage Factor. The gage factor can be entered as a constant used for all
repetitions or a variable array can be loaded with individual gage factors
which are automatically used with each rep. To use an array enter the
parameter as arrayname(), with no element number in the parentheses.
v
Poisson ratio, enter 0 if it does not apply to configuration.
StrainCalc Example
This example uses StrainCalc to find the microstrain value of a bridge output.
'
Program name: STRAIN.DLD
Public Count, ZStrain, StMeas, Strain, Flag(8)
' Declare all variables as
public
'Data Table STRAINS samples every measurement when user Sets Flag(1)
High
DataTable(STRAINS,Flag(1),-1)
DataInterval(0,0,0,100)
'Interval = Scan, 100 lapses
Sample (1,Strain,Ieee4)
EndTable
'DataTable ZERO_1 stores the "zero" measurements
DataTable(ZERO_1,Count>99,100)
'Trigger on Count 100
Average(1,ZStrain,IEEE4,0)
EndTable
'Subroutine to measure Zero, Called on first pass or when user sets Flag(2)low
Summary of Contents for CR5000
Page 40: ...Section 1 Installation and Maintenance 1 16 This is a blank page ...
Page 54: ...Section 2 Data Storage and Retrieval 2 14 This is a blank page ...
Page 88: ...Section 4 CRBasic Native Language Programming 4 12 This is a blank page ...
Page 94: ...Section 5 Program Declarations 5 6 This is a blank page ...
Page 172: ...Section 7 Measurement Instructions 7 46 This is a blank page ...
Page 206: ......
Page 236: ...Index Index 4 This is a blank page ...
Page 237: ...This is a blank page ...