CPEC300/306/310 Closed-Path Eddy-Covariance Systems
24
5. Installation
The following tools are required to install a CPEC300-series system in the
field. Additional tools may be required for a user-supplied tripod or tower.
•
9/16-in, open-end wrench
•
1/2-in, open-end wrench
•
11/16-in, open-end wrench
•
Adjustable wrench
•
Small, flat-tip screwdriver (included with EC100 and CPEC300-series
system)
•
Large, flat-tip screwdriver
•
Sledgehammer (to drive grounding rod into the ground)
•
3/16-in hex-key wrench (included with CM250 leveling mount)
5.1 Mounting
5.1.1 Support Structure
A CPEC300-series system has three major components that must be mounted
to a user-provided support structure.
EC sensors (EC155 and CSAT3A):
Mounted on a horizontal round pipe of
3.33 cm (1.31 in) outer diameter, such as the CM20X crossarm as in FIGURE
EC100 electronics (denoted as “CPEC300 Closed-Path Eddy-Covariance
System” for the CPEC300):
Mounted within 3.0 m (10.0 ft) of the EC
sensors. The EC100 mounting bracket will accommodate a pipe at any
orientation, with outer diameter from 2.5 cm to 4.8 cm (1.0 in to 1.9 in).
For CPEC 300:
CPE300 pump module enclosure:
Mounted within 3.0 m (10.0 ft) of the
CPEC300 enclosure. The pump module enclosure is similar to the ENC10/12,
with the same mounting options (tower, tripod, leg, or pole).
For CPEC306/310:
CPEC306 or CPEC310 enclosure:
Mounted where it can be accessed easily
to retrieve data from the microSD cards in the datalogger. The CPEC306 or
310 enclosure is similar to the ENC16/18, with the same mounting options
(tower, tripod, leg, or pole).
The following sections describe a typical application using a CM210 tripod and
CM202 crossarm. The CM210 tripod and leg mounting options are ideal for a
low EC measurement height to minimize wind disturbance.
5.1.2 Mount Enclosures
Mount the EC100 electronics within 3.0 m (10.0 ft) of the EC sensors (this
measurement corresponds to the length of the cables on the EC155 and the
CSAT3A).
For the EC100 and the system enclosure, open the sealed bag containing the
desiccant packs and humidity card. Place two of the desiccant packs and the