Appendix H. Output Variables
H-8
•
ERROR: Cannot run Zero/Span sequence - CHECK_ZERO is False
•
ERROR: Cannot run Zero/Span sequence - Check valve temperature
•
ERROR: Cannot run Zero/Span - Check scrub module temperature
If a CPEC300 system begins the startup process for a zero/span sequence, and
then detects an error, one of the following messages will be displayed:
•
Zero/Span startup manually aborted by turning pump on
•
ERROR: Zero/Span startup aborted - Check scrub module
temperature
•
ERROR: Zero/Span startup aborted - Check valve temperature
If a CPEC300 system has already started the zero/span sequence and then
detects some error condition that causes the zero/span sequence to abort, one of
the following messages will be displayed:
•
Zero/Span sequence manually aborted by turning pump on
•
ERROR: Zero/Span sequence aborted - Check scrub module
temperature
•
ERROR: Zero/Span sequence aborted - Check valve temperature
If a CPEC300 system attempts to set the IRGA zero or span as part of a
zero/span sequence and there is a problem identified (
diag_cpec
is nonzero),
the CPEC300 system will skip the zero or span setting step, continue the rest of
the sequence, and display one of the following messages. If the user attempts to
set the zero or span manually this error or check is bypassed, and the zero or
span will be set regardless of the status of
diag_cpec
.
•
ERROR: Automatic Zero aborted - Check diagnostics
•
ERROR: Automatic CO2 Span aborted - Check diagnostics
Some of the messages relate to the use of the valve module in manual mode:
•
Valve operation not allowed when pump is on
•
EC mode restarted by turning pump on
•
ERROR: Valves disabled - Check valve temperature
•
ERROR: Valves disabled - Check press_offset
•
ERROR: Valves disabled - Check scrub module temperature
Some of the messages relate to the configuration of the EC155:
•
ERROR: Powering OFF EC155 gas head
•
Powering ON EC155 gas head
•
Configuring the EC100
•
Setting the zero
•
Setting the CO2 span
•
Setting the H2O span