Appendix G. Public Variables
G-2
TABLE G-1. CPEC300 public variables
When Defined
Usage
Variable Name
Units
Description
always
SONIC
diag_sonic
CSAT3A diagnostic word
always
IRGA
CO2
µmol∙mol
-1
CO
2
mixing ratio, relative to
dry air
always
IRGA
H2O
µmol∙mol
-1
H
2
O vapor mixing ratio,
relative to dry air
always
IRGA
diag_irga
EC155 diagnostic word
always
IRGA
cell_tmpr
ºC
Temperature of the sample cell
always
IRGA
cell_press
kPa
pressure in the sample cell
always
IRGA
CO2_signal
Relative strength of the infrared
signal in the CO
2
absorption
band
always
IRGA
H2O_signal
Relative strength of the infrared
signal in the H
2
O absorption
band
always
IRGA
diff_press
kPa
Differential pressure (sample
cell, relative to ambient)
always
IRGA
source_tmpr
ºC
Temperature measured inside
the source housing
always
IRGA
not_used
Placeholder variable to receive
a value from the EC100 (used
for an EC150, but not used for
an EC155)
always
INFO
Td_cell
ºC
H
2
O converted to dewpoint
temperature in the sample cell
always
INFO
Td_ambient
ºC
Dewpoint temperature
corrected for pressure
difference between the sample
cell and ambient
always
USER
pump_ON
set = True to enable the sample
pump
always
INFO
pump_flow
LPM
sample pump speed, converted
to volumetric flow rate
always
INFO
pump_control
pump speed control variable;
can be from 0 (off) to 1 (full
speed)
always
INFO
pump_press
kPa
pressure measured at the
sample pump inlet
always
INFO
pump_tmpr
ºC
temperature of the pump
always
DIAG
pump_tmpr_OK
Boolean flag: True if the pump
temperature is within its
operating range
always
INFO
pump_heat_ON
Boolean flag: True if the pump
heater is on
always
INFO
pump_fan_ON
Boolean flag: True if the pump
fan is on
VALVE_MODULE
USER
STARTsequence
Set = True to manually initiate
a zero/span sequence
VALVE_MODULE
INFO/USER valve_number
Valve number (0 to 6);
determines which zero/span
valve is selected