CMP6, CMP10, CMP11, and CMP21 Pyranometers
6
TABLE 6-1. CMP-series Specifications
Specification
CMP6
CMP10/CMP11
CMP21
ISO Classification
First Class
Secondary Standard
Maximum irradiance
2000 W
•
m
–2
4000 W
•
m
–2
Spectral range
(50% points)
285 to 2800 nm
Response time (95 %)
<18 s
<5 s
Expected daily
uncertainty
<5%
<2%
Zero offset due to
thermal radiation
(200 W
•
m
–2
)
<15 W
•
m
–2
<7 W
•
m
–2
Zero offset due to
temperature change
(5 K
•
hr
–1
)
<4 W
•
m
–2
<2 W
•
m
–2
Non-stability
(change/year)
<1 %
<0.5%
Non-linearity
(0 to 1000 W
•
m
–2
)
<1%
<0.2%
Directional error
(up to 80° with 1000
W
•
m
–2
beam)
<20 W
•
m
–2
<10 W
•
m
–2
Tilt error
(at 1000 W
•
m
–2
)
<1%
<0.2%
Level accuracy
0.1°
Operating temperature
–40 to 80 °C
Temperature dependence
of sensitivity
<4% (–10 to 40 °C)
<1% (–10 to 40 °C)
<1% (–20 to 50 °C)
Sensitivity
5 to 20 µV / W
•
m
–2
7 to 14 µV / W
•
m
–2
Typical signal output for
atmospheric applications
0 to 20 mV
0 to 15 mV
Weight
0.6 kg (1.3 lb) without cable; 0.9 kg (2 lb) with 10 m (33 ft) cable
Impedance
1
20 to 200
Ω
10 to 100
Ω
1
Impedance is defined as the total electrical impedance at the radiometer output connector fitted to the housing. It arises
from the electrical resistance in the thermal junctions, wires, and passive electronics within the radiometer.
EU Declaration of
Conformity:
View at:
6.2 CVF4 Ventilation Unit
Compatible pyranometers:
CMP6, CMP10, CMP11, CMP21
Power supply:
12 VDC, 0.9 A (with 5.5 W heater)
Operating temperature range:
–40 to 70 °C
Ventilation power:
5 W continuously