CM3 Pyranometer
7
Example 1
;{CR10X}
;
*Table 1 Program
01: 10
Execution Interval (seconds)
1: Volt (Diff) (P2)
1:
1
Reps
2:
24
250 mV 60 Hz Rejection Range
3:
1
DIFF Channel
4:
1
Loc [ W_m2 ]
5:
45.725
Mult ;multiplier = (1 / 0.02187 mV / W/m2)
6:
0.0
Offset
;Set negative values to zero.
;
2: If (X<=>F) (P89)
1:
1
X Loc [ W_m2 ]
2:
4
<
3:
0
F
4:
30
Then Do
3: Z=F x 10^n (P30)
1:
0.0
F
2:
00
n, Exponent of 10
3:
1
Z Loc [ W_m2 ]
4: End (P95)
5: If time is (P92)
1:
0
Minutes (Seconds --) into a
2:
60
Interval (same units as above)
3:
10
Set Output Flag High (Flag 0)
6: Real Time (P77)
1:
1220
Year,Day,Hour/Minute (midnight = 2400)
7: Average (P71)
1:
1
Reps
2:
1
Loc [ W_m2 ]
*Table 2 Program
02: 0.0000
Execution Interval (seconds)
*Table 3 Subroutines
End Program
5.2 Total Solar Radiation
In Example 2 a CR10X is used to record daily total flux density. This total flux
density is in MJ m
-2
day
-1
to avoid the need for high resolution discussed in
5.2.1. Negative values are set to zero before they are added to the running
total.