Campbell 237 Product Manual Download Page 13

237 Leaf Wetness Sensor 

TABLE 7-2.  Measurement Instructions, Parameters, Results 

Datalogger 

Measurement 

Instruction 

Excitation 

(mV) 

Input 

Range 

Code 

Integration/ 

f

N1

 

Multiplier  Offset  Result 

CR6, 

CR1000X 

BrHalf () 

2500 

mV200 or 

mV200C

1

 

15000 

Vs/Vx 

CR300 

BrHalf () 

2500 

mV2500 

4000 Hz 

Vs/Vx 

CR200(X) 

ExDelSE () 

2500 

n/a 

500 µs 

0.0004 

Vs/Vx 

CR800, 

CR1000 

BrHalf () 

2500 

mV25 or 

mV25C

1

 

250 µs 

Vs/Vx 

CR3000 

BrHalf () 

5000 

mV50 or 

mV50C

1

 

250 µs 

Vs/Vx 

1

 The mV200C, mV25C, and mV50C input range codes check for an open connection indicating that no sensor is connected 

or a cable has been cleanly cut. If an open connection occurs, NAN will be returned. A floating voltage can occur that 

falsely indicates the presence of a missing sensor when the mV200C, mV25C, and mV50C input range codes are not used. 

7.4.2  Calculating Sensor Resistance 

The CRBasic program should include an expression that calculates sensor 

resistance. With reference to FIGURE 

7-2

, sensor resistance (Rs), expressed in 

kΩ, is calculated as follows:

  

 

Rs = R

2

 / (Vs/Vx) – R

2

 – R

1

 

Therefore, 

Rs (kΩ) = 1/(Vs/Vx) 

– 101 

Except for the CR200(X), Campbell Scientific also suggests including the 

following after the resistance calculation: 

If Rs > 10000 Then 
  Rs = 10000 
EndIf 

The high resolution of Campbell Scientific dataloggers can pick up very small 

signals, which can result in an erratic signal when the sensor is dry. The 

wet/dry transition is less than 1000 

, and therefore, 10,000 

kΩ 

gives plenty 

of range outside of what one would consider to be dry. 

8. Operation

An exact range of measurements is impossible to give since the 237 is field 

calibrated. The sensor is not precise and the quality of water bridging the 

electrodes varies. As demonstrated in Appendix 

B

Example Programs

 

(p. B-1)

, a 

common practice is to measure grid resistance in 

kΩ

 using a one bin histogram 

to calculate at what fraction of the output interval the sensor is wet (Section 

8.1

Calculating Wet Time Fraction

 

(p. 8)

). If resistance meets the threshold 

(

≤150 kΩ

 in the example program), the grid is considered wet. Since the output 

interval is 60 minutes, if the histogram fraction equals 0.33, the leaf was wet 

for 20 minutes during that hour. 

Summary of Contents for 237

Page 1: ...237 Leaf Wetness Sensor Revision 01 2021 Copyright 1988 2021 Campbell Scientific CSL I D 506...

Page 2: ...under guarantee or not Please state the faults as clearly as possible and if the product is out of the guarantee period it should be accompanied by a purchase order Quotations for repairs can be give...

Page 3: ...cell phones and aerials may also not be applicable according to your locality Some brackets shields and enclosure options including wiring are not sold as standard items in the European market in some...

Page 4: ...beginning work Wear a hardhat and eye protection and take other appropriate safety precautions while working on or around tripods and towers Do not climb tripods or towers at any time and prohibit cl...

Page 5: ...nsor Preparation 4 7 2 Sensor Deployment 4 7 3 Wiring to Datalogger 5 7 4 Datalogger Programming 6 7 4 1 Measurement of Vs Vx 6 7 4 2 Calculating Sensor Resistance 7 8 Operation 7 8 1 Calculating Wet...

Page 6: ...Table of Contents ii 7 2 Measurement Instructions Parameters Results 7 9 1 Symptom Causes 9 CRBasic Examples B 1 CR1000 Program Example B 1 B 2 CR200 X Program Example B 1...

Page 7: ...rubber This compound was chosen for its resistance to temperature extremes moisture and UV degradation However this jacket will support combustion in air It is rated as slow burning when tested accor...

Page 8: ...ocessing will be required to determine parameters such as the percent of time wet Section 8 1 Calculating Wet Time Fraction p 8 4 After selecting the sensor click Wiring Diagram to see how the sensor...

Page 9: ...Scientific supplies the sensor unpainted because individual modifications will vary with the application The paper listed in Section 10 References p 9 describes the effects of paint colour and sensor...

Page 10: ...in plant disease and other applications Besides providing some protection for the gold plated electrodes this paint allows tiny water droplets to spread and bridge the electrodes Gillespie and Kidd1...

Page 11: ...uit schematic of the 237 TABLE 7 1 describes wiring to Campbell Scientific dataloggers FIGURE 7 2 237 Sensor Schematic TABLE 7 1 Wire Colour Function and Datalogger Connection Wire Colour Wire Functio...

Page 12: ...ndix A Importing Short Cut Code Into CRBasic Editor p A 1 Programming basics for CRBasic dataloggers are in the following section Complete program examples for select CRBasic dataloggers can be found...

Page 13: ...pressed in k is calculated as follows Rs R2 Vs Vx R2 R1 Therefore Rs k 1 Vs Vx 101 Except for the CR200 X Campbell Scientific also suggests including the following after the resistance calculation If...

Page 14: ...er to Appendix B Example Programs p B 1 for information on calculating fraction of time wet with the CR200 X 8 2 Calibration A wet dry threshold of 150 k is used in Appendix B Example Programs p B 1 W...

Page 15: ...ly cut if the mV200C mV25C or mV50C range code was used for the BrHalf instruction Another cause of these outputs is a highly conductive state 25 mV that occurs if the sensor is very wet with very ion...

Page 16: ...can now be edited for additional refinement Once the file is edited with CRBasic Editor Short Cut can no longer be used to edit the program it created 2 To add the Short Cut wiring information into t...

Page 17: ...V25 1 VX1 1 2500 True 0 250 1 0 Rs_kOhms 1 Vs_Vx 101 If Rs_kOhms 10000 Then Rs_kOhms 10000 EndIf CallTable Wetness NextScan EndProg CRBasic Example B 2 CR200 X Program Example CR200 X Series Datalogge...

Page 18: ...lculate Time Fraction Wet at top of each hour If IfTime 0 60 Min Then Interval must match data table interval TimeFractionWet ScanIntervalWet ScanIntervalSum ScanIntervalWet 0 ScanIntervalSum 0 EndIf...

Page 19: ...nfo campbellsci fr www campbellsci fr Germany Location Phone Email Website Bremen Germany 49 0 421 460974 0 info campbellsci de www campbellsci de India Location Phone Email Website New Delhi DL India...

Reviews: