PREPARATION FOR WASHING.
Effective washing starts with preparation of the glassware. Ensure that any solid residue is
removed, if necessary by rinsing or pre-soaking (in a non-foaming solution) before carefully placing
the glassware on the appropriate basket tray or jet rack. Avoid overloading basket trays and jet
racks.
Avoid long delays between loading the machine and using it. Deposits could dry and harden onto
the glassware or chemicals could drip from dirty glassware into the wash chamber and could
cause corrosion or staining of the stainless steel.
Items must be placed on an appropriate basket tray or jet rack. Narrow mouthed glassware may
not be washed internally if placed on a basket tray as the wash water may not be sprayed inside
effectively. Such items should be washed on jet racks, so that an adequate supply of wash water
can reach all the inside surfaces.