5.8.3 Forced resin regeneration
Press
ESCAPE.
When prompted enter
superuser password
.
PROGRAMME
screen is shown.
Scroll to
UTILITIES
, press
ENTER
. Scroll to
REGENERATION,
press
ENTER
.
Scroll down to
LT WATER
and enter
5
in the menu below using the
PLUS
key and confirm with
ENTER
key. The display will ”
update
”.
HARDNESS IN °F <00.0>
HARDNESS IN °T <00.0>
HARDNESS IN °I <00.0>
LT WATER <00.5>
ESCAPE
out to select
Prog 1
, (Prewash) and press
START
. At the end, after the prewash water
has been pumped out, the display will show “
REGENERATION
“ whilst regeneration takes place.
Afterwards, re-enter the regeneration menu and input the correct value onto French Hardness (top
line, typically
30
). Press
ENTER
to update, then
ESCAPE
back to programs.
At the start of the next chosen program, there will be several minutes of “
Resin Flush
” before the
“Prewash” phase of the program commences.