Camille Bauer SINEAX AM3000 Device Handbook Download Page 86

PM 1001484 000 08  

Device handbook SINEAX DM5000

 

 

  86/96

 

B2  Display matrices for split-phase (two-phase) systems

 

 

Display menu 

Corresponding matrix 

 

 

 

U1N 

U1N_MM 

 

 

U2N 

U2N_MM 

 

 

U_MM 

 

 

UNE 

UNE_MAX 

 

 

I1 

I1_MAX 

I2 

I2_MAX 

IN 

IN_MAX 

IPE 

IPE_MAX 

P1 

P_MAX 

P1_MAX 

P2 

Q_MAX 

P2_MAX 

Q1 

S_MAX  

Q1_MAX 

PF 

Q2 

F_MM 

Q2_MAX 

P_TRIANGLE 

P1_TRIANGLE 

P2_TRIANGLE 

 

 

 

PF_MIN 

Cφ_MIN

 

 

 

 

 

I> 1.1 / 1.2 

I> 2.1 / 2.2 

 

 

ϑ

 1.1 / 1.2 

ϑ

 2.1 / 2.2 

 

 

 

 

 

 

 

 

ΣP_I_IV_HT

 

ΣP_I_IV_NT

 

ΣP_II_III _NT

 

ΣP_II_III_HT

 

ΣQ_I_II_HT

 

ΣQ_I_II _NT

 

ΣQ_III_IV _HT

 

ΣQ_I_II _NT

 

 

 

 

 

 

 

ΣMETER1

 

ΣMETER2

 

ΣMETER3

 

ΣMETER4

 

ΣMETER5

 

ΣMETER6

 

ΣMETER7

 

ΣMETER8

 

ΣMETER9

 

ΣMETER10

 

ΣMETER11

 

ΣMETER12

 

 

 

 

 

 

 

MT_P_I_IV 

MT_P_II_III 

MT_Q_I_II 

MT_Q_III_IV 

MT_S 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

M1   

TR_M1   

M1_MM 

 

 

M2   

TR_M2   

M2_MM 

 

 

M3  

TR_M3   

M3_MM 

 

 

M4   

TR_M4   

M4_MM 

 

 

M5  

TR_M5   

M5_MM 

M6   

TR_M6   

M6_MM 

M7   

TR_M7   

M7_MM 

M8  

TR_M8   

M8_MM 

M9  

TR_M9   

M9_MM 

M10   

TR_M10   

M10_MM 

M11 

TR_M11 

M11_MM 

M12  

TR_M12 

M12_MM 

 

 

 

 

 

 

IB1 
IB2 
IB1_MAX 
IB2_MAX 

 

 

 

 

Summary of Contents for SINEAX AM3000

Page 1: ...X DM5000 Operating Instructions SINEAX DM5000 2019 06 Camille Bauer Metrawatt AG Aargauerstrasse 7 CH 5610 Wohlen Switzerland Phone 41 56 618 21 11 Telefax 41 56 618 35 35 E Mail info cbmag com https www camillebauer com Legal information ...

Page 2: ...le to understand and follow the given safety and warning notices Intended use The product described in this document may be used only for the application specified The maximum electrical supply data and ambient conditions specified in the technical data section must be adhered For the perfect and safe operation of the device proper transport and storage as well as professional assembly installatio...

Page 3: ... 27 5 14 Uninterruptible power supply UPS 27 5 15 GPS time synchronization 28 6 Commissioning 30 6 1 Operating LED 30 6 2 Parametrization of the device functionality 31 6 3 Installation check 31 6 4 Ethernet installation 33 6 4 1 Settings 33 6 4 2 Connection of the standard interface 35 6 4 3 Connection of the IEC61850 interface 36 6 4 4 Connection of the PROFINET interface 36 6 4 5 MAC addresses ...

Page 4: ...stment 64 8 2 Cleaning 64 8 3 Battery 64 8 4 Disposal 64 9 Technical data 65 Annex 72 A Description of measured quantities 72 A1 Basic measurements 72 A2 Harmonic analysis 76 A3 System imbalance 77 A4 Mean values and trend 78 A5 Meters 79 B Display matrices 80 B0 Used abbreviations for the measurements 80 B1 Display matrices for single phase system 85 B2 Display matrices for split phase two phase ...

Page 5: ...the field of electrical engineering is required For assembly and installation of the device knowledge of applicable national safety regulations and installation standard is required 1 2 Scope of supply Measurement device SINEAX DM5000 Safety instructions multiple languages Battery pack optional for devices with UPS only 1 3 Further documents The following documents are provided electronically via ...

Page 6: ...versal measurement and monitoring in power systems A full parameterization of all functions of the device is possible directly at the device for versions with display or via web browser The universal measurement system of the device may be used directly for any power system from single phase up to 4 wire unbalanced networks without hardware modifications Using additional optional components the op...

Page 7: ...at dangerous connections of the device can not be touched during operation and that the spread of flames heat and smoke from the interior is prevented This may be done by providing an enclosure e g case cabinet or using a room accessible to qualified personal only and compliant with local fire safety standards The standard version of the DM5000 can be clipped onto a top hat rail according to EN607...

Page 8: ... nominal voltages up to 1000 V Nameplate of a device with TFT display Ethernet interface Modbus RTU interface Data logger 2 relay outputs 4 analog outputs UPS Symbol Meaning Device may only be disposed of in a professional manner Double insulation device of protection class 2 CE conformity mark The device fulfills the requirements of the applicable EU directives Products with this mark comply with...

Page 9: ... X1 1 X1 3 2 1 X4 2 X4 3 2 2 X4 1 X4 3 2 3 X3 2 X3 3 2 4 X3 1 X3 3 4 digital inputs active 1 1 X1 1 X1 3 1 2 X1 2 X1 3 1 3 X2 1 X2 3 1 4 X2 2 X2 3 2 1 X3 1 X3 3 2 2 X3 2 X3 3 2 3 X4 1 X4 3 2 4 X4 2 X4 3 4 digital inputs passive 1 1 X1 1 X1 3 1 2 X1 2 X1 3 1 3 X2 1 X2 3 1 4 X2 2 X2 3 2 1 X3 1 X3 3 2 2 X3 2 X3 3 2 3 X4 1 X4 3 2 4 X4 2 X4 3 2 temperature inputs 1 1 X1 2 X1 3 1 2 X2 2 X2 3 2 1 X3 2 X3...

Page 10: ... must originate at circuit breakers or fuses rated 5 Amps or less This does not apply to the neutral connector You have to provide a method for manually removing power from the device such as a clearly labeled circuit breaker or a fused disconnect switch in accordance with IEC 60947 2 or IEC 60947 3 When using voltage transformers you have to ensure that their secondary connections never will be s...

Page 11: ... transformer can be omitted If voltage UNE does not need to be measured connection of PE can be omitted With current and voltage transformer If current IN or voltage UNE does not need to be measured the corresponding transformers can be omitted L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 N PE 5 A UL listed L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 N K k L l 5 A UL l...

Page 12: ...voltage transformers In case of current measurement via L2 or L3 connect the device according to the following table Terminals 1 3 2 5 8 Current meas via L2 I2 k I2 l L2 L3 Current meas via L3 I3 k I3 l L3 L1 L1 L2 L3 L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 5 A UL listed L1 L2 L3 K k L l L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 5 A UL listed L1 L2 L3 K k L l L1 L2 L3...

Page 13: ...voltage transformers In case of current measurement via L2 or L3 connect the device according to the following table Terminals 1 3 2 5 8 Current meas via L2 I2 k I2 l L3 L1 Current meas via L3 I3 k I3 l L1 L2 L1 L2 L3 L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 5 A UL listed L1 L2 L3 K k L l L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 5 A UL listed L1 L2 L3 K k L l L1 L2 L3...

Page 14: ...voltage transformers In case of current measurement via L2 or L3 connect the device according to the following table Terminals 1 3 2 5 8 Current meas via L2 I2 k I2 l L2 L1 Current meas via L3 I3 k I3 l L3 L2 L1 L2 L3 L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 5 A UL listed L1 L2 L3 K k L l L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 5 A UL listed L1 L2 L3 K k L l L1 L2 L3...

Page 15: ...he device according to the following table Terminals 1 3 2 5 8 Current meas via L2 I2 k I2 l L2 L3 L1 Current meas via L3 I3 k I3 l L3 L1 L2 By rotating the voltage connections the measurements U12 U23 and U31 will be assigned interchanged L1 L2 L3 L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 5 A UL listed L1 L2 L3 K k L l L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 5 A UL l...

Page 16: ...oltage transformer If voltage UNE does not need to be measured the corresponding transformer can be omitted In case of current measurement via L2 or L3 connect the device according to the following table Terminals 1 3 2 11 Current meas via L2 I2 k I2 l L2 N Current meas via L3 I3 k I3 l L3 N L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 N PE 5 A UL listed L1 L2 L3 N I1 3 1 I2 4 ...

Page 17: ...ransformers With current and 3 single pole isolated voltage transformers L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 5 A UL listed L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 K k L l K k L l K k L l 5 A UL listed L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 K k L l K k L l K k L l x u X U X U X U x u x u 5 A UL listed ...

Page 18: ...h With current transformers With current and 3 single pole isolated voltage transformers L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 5 A UL listed L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 K k L l K k L l 5 A UL listed L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 K k L l K k L l x u X U X U X U x u x u 5 A UL listed ...

Page 19: ...itted If current IN does not need to be measured the corresponding transformer can be omitted With current and voltage transformer If current IN does not need to be measured the corresponding transformer can be omitted L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 N 5 A UL listed PE L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 N K k L l K k L l K k L l 5 A UL...

Page 20: ...oes not need to be measured the corresponding transformer can be omitted With current and voltage transformer If current IN does not need to be measured the corresponding transformer can be omitted L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 N 5 A UL listed PE L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 L3 N K k L l K k L l K k L l 5 A UL listed K k L l PE L1...

Page 21: ...transformer In systems without a primary neutral conductor a voltage transformer with a secondary center tap can also be used L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 N 5 A UL listed PE L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 N K k L l K k L l 5 A UL listed K k L l PE L1 L2 L3 N I1 3 1 I2 4 I3 9 7 IN PE 6 10 12 16 11 2 5 8 L1 L2 N K k L l K k L l 5 A UL l...

Page 22: ...ng I O extensions only 5 8 Digital inputs The device provides a standard passive digital input In addition depending on the device version there may be 4 channel passive or active digital input modules available Usage of the standard digital input Status input Meter tariff switching Usage of the inputs of the optional input modules Counting input for pulses of meters for any kind of energy pulse w...

Page 23: ...Recommended if input impedance of counter 100 kΩ Driving a counter mechanism The width of the energy pulses can be selected within a range of 30 up to 250ms but have to be adapted to the external counter mechanism Electro mechanical meters typically need a pulse width of 50 100ms Electronic meters are partly capable to detect pulses in the kHz range There are two types NPN active negative edge and...

Page 24: ...system to connect 5 11 Modbus interface RS485 Via the optional Modbus interface measurement data may be provided for a superior system However the Modbus interface cannot be used for device parameterization 1 One ground connection only This is possibly made within the master PC Rt Termination resistors 120 Ω each for long cables approx 10 m Rs Bus supply resistors 390 Ω each The signal wires A B h...

Page 25: ... Meas transformer Residual current transformer 500 1 up to 1000 1A Rated burden 100 Ω 0 025 VA up to 200 Ω 0 06 VA Use only transformers intended for this application according to our current transformer catalog or transformers that fulfill the above specification Using transformers with divergent specifications may damage the measurement inputs Connection The current transformers including the co...

Page 26: ...ld be mounted or installed close to strong magnetic fields Measurement lines should also not be laid in parallel to power lines 6 For measurement range 1A only The rated output of the transformer must be chosen that it is reached when the rated secondary current 1A flows Consider that the burden of the transformer is not only made up by the burden of the measurement input but also by the resistanc...

Page 27: ...pack for the uninterruptible power supply is supplied separately Please note that compared to the storage temperature range of the base unit the storage temperature range of the battery pack is restricted Ensure that devices with uninterruptible power supply are used in an environment in accordance with the specification Outside this operating temperature range it is not ensured that the battery p...

Page 28: ...1Hz pulse accuracy 1μs Connector RJ45 Choosing a mounting location For a correct operation the GPS receiver requires data from at least 3 satellites at the same time Therefore position the receiver so that the clearest possible view of the sky and horizon in all direction is obtained This can be on the roof of a building at best without reception being restricted by other buildings or obstacles Av...

Page 29: ...ing or sharp kinking of the cable should be avoided Commissioning In the settings menu change time synchronization to NTP server GPS Check the time synchronization status The time synchronization can be restarted by switching the time synchronization off and on again Time synchronization via GPS and NTP server may work in parallel If both synchronization sources are available the system uses the m...

Page 30: ...quency 1 Serial number 2 Test and conformity marks 3 Assignment voltage inputs 4 Assignment current inputs 5 Assignment power supply 6 Load capacity relay outputs 6 1 Operating LED The operating LED shows the present device state Procedure LED display Booting of device Flashes green 1 Hz If successful Change to static green display Firmware update Change to update mode Static red During update Fla...

Page 31: ...he current and voltage inputs can be checked in two ways a Sense of rotation check Using the sequence of the current and voltage phasors the sense of rotation is determined and compared to the configured one The phase rotation indicator is arranged in the menu Phasor diagram Test requirement Magnitude of all connected voltages at least 5 of nominal magnitude of all connected currents at least 0 2 ...

Page 32: ...polarity of current I2 What s wrong Voltage sequence L1 L3 L2 L3 and L2 seems to be interchanged Current sequence L1 L2 L3 Angle U I Angles between UL2 IL2 and UL3 IL3 do not correspond to the expectations Required correction Exchanging the connections of the voltages L2 and L3 What s wrong Voltage sequence L1 L3 L2 L3 and L2 seems to be exchanged Current sequence L1 L3 L2 Current L2 is out of the...

Page 33: ...with the network administrator IP address This one must be unique i e may be assigned in the network only once Subnet mask Defines how many devices are directly addressable in the network This setting is equal for all the devices Examples Gateway address Is used to resolve addresses during communication between different networks It should contain a valid address within the directly addressable ne...

Page 34: ...t mask 255 255 252 0 11111111 11111111 11111100 00000000 variable range xx xxxxxxxx First address 192 168 56 0 11000000 10101000 00111000 00000000 Last address 192 168 59 255 11000000 10101000 00111011 11111111 The device 192 168 57 64 can access directly the devices 192 168 56 0 192 168 59 255 DHCP If a DHCP server is available alternatively the mode DHCP or DHCP addresses only can be selected fo...

Page 35: ...ning the type of communication As a standard Modbus TCP communication is performed via TCP port 502 NTP uses port 123 However the port for the Modbus TCP communication may be modified You may provide a unique port to each of the devices e g 503 504 505 etc for an easier analysis of the communication traffic Independent of these setting a communication via port 502 is always supported The device al...

Page 36: ...connecting Ethernet cables Both ports are equivalent and internally connected via a switch Note The interface may only be connected to a local Profinet network which is designed as SELV circuit according to IEC 60950 1 Interface RJ45 sockets Ethernet 100BaseTX Mode 10 100 MBit s full half duplex auto negotiation Protocols PROFINET LLDP SNMP Functionality of the LED s LED State Meaning X1 green X2 ...

Page 37: ...n on the nameplate Port connector X1 Chassis MAC 2 Port connector X2 Chassis MAC 1 6 4 6 Communication tests Via the service menu on the device website you may check if the selected network structure is valid The device must be able to reach the DNS server via gateway The DNS server then allows resolving the URL of the NTP server to an IP address The Standard Ethernet interface serves as interface...

Page 38: ...www camillebauer com dm5000 en 6 6 PROFINET IO interface The PROFINET interface provides a cyclical process image which can be freely assembled by the user 6 6 1 General stations description file GSD The GSD file describes the functionality available via the PROFINET interface of the device During system design by means of a configuration tool e g TIA or Simatic Step 7 of Siemens the GSD file serv...

Page 39: ... the device As soon as the GSD file has been imported the device is available in the hardware catalog and can be integrated using drag drop There are three models available that represent the different designs of the whole device series The selection shown above is for example suited for the devices AM2000 AM300 and PQ3000 which have the same design panel 144x144mm and support the same measured va...

Page 40: ...her details are not given here Assembly of the cyclical process image In Slot 1 always the module System state is present providing the following information Bit Meaning 0 0 Measurement system stopped or not reachable 1 Measurement system running 1 0 1 When the measurement system is running the bit changes its state when the value of at least one of the modules changes 2 31 not used currently set ...

Page 41: ...alanced The Modbus device description provides the information about the validity of the measurements with respect to the used system configuration This description can be downloaded via one of the following sources Homepage https www camillebauer com dm5000 en USB stick with software and documentation no 156 027 optional If invalid measurements are used in the process image their values are alway...

Page 42: ...ge 6 8 Security system 6 8 1 Protection against device data changing Configuration or measurement data stored in the device may be modified via either service or settings menu To protect these data a security system may be activated default not activated If the security system is active the user has to enter a password before executing protected functions Subsequent to a successful password input ...

Page 43: ...ither use a Camille Bauer certificate or its own customer certificate This may be selected when activating https communication via the Settings of the Security system in the item Web Security Camille Bauer certificate The certificate is provided via our webpage https camillebauer com dm5000 en Once the certificate is downloaded to the local computer the certificate can be installed manually Just d...

Page 44: ...fine an IP address list of up to 10 clients allowed to have access to the device All other clients will be blocked Enable the whitelist via the Settings of the Security system in the item Whitelist If a DHCP server is used in the system clients may get different IP addresses on each startup losing this way access to the device If a device is no longer accessible you can reset its IP address LAN de...

Page 45: ...id functionality of the keys is then shown in a help bar 7 2 Selecting the information to display For devices with display information selection is performed via menu Menu items may contain further sub menus Displaying the menu Press ESC Each time the key is pressed a change to a higher menu level is performed if present Displaying information The menu item chosen using can be selected using OK Re...

Page 46: ...ent displays and used symbols For displaying measurement information the device uses both numerical and numerical graphical measurement displays Examples Measurement information 2 measured quantities 4 measured quantities Graphical measurement display Further examples ...

Page 47: ...bination of them or an appropriate graphical representation is used You can select your own point of view by selecting the reference arrow system load or generator in the settings of the measurement Used symbols For defining a measurement uniquely a short description e g U1N and a unit e g V are often not sufficient Some measurements need further information which is given by one of the following ...

Page 48: ...system etc Hints U I transformer The primary to secondary ratio is used only for converting the measured secondary to primary values so e g 100 5 is equivalent to 20 1 The values do not have any influence on the display format of the measurements Nominal voltage current Used only as reference values e g for scaling the harmonic content TDD of the currents Maximum primary values U I These values ar...

Page 49: ...used for triggering the summary alarm and selection of a possible source for resetting Operating hours Selection of the running condition for up to 3 operating hour counters Digital outputs Digital output State pulse or remote controlled digital output with source pulse width polarity number of pulses per unit Digital outputs Relay State or remote controlled relay output with source Analog outputs...

Page 50: ...vice homepage via http ip_addr The default IP address of the device is 192 168 1 101 This request works only if device and PC are in the same network when applying the subnet mask examples Via WEB GUI all device settings can be performed as via the local GUI Possibly modifications needs to be saved in the device before all parameters have been set In such a case the following message appears If th...

Page 51: ...ded into the device The values in the WEB GUI will be updated accordingly Normally devices differ in the settings of network or Modbus parameters and device name Thus when loading the file you can choose whether the appropriate settings of the device should be retained or overwritten by the values in the file to be uploaded Storing the current parameter settings of the WEB GUI into the device Savi...

Page 52: ...mains active until the associated measured quantity falls below the limit for OFF state again The limit value is inactive 0 if either the limit for ON is not yet reached or if following the activation of the limit value the associated measured quantity falls below the limit for OFF state again Lower limit Limit for ON Limit for OFF The limit value becomes active 1 as soon as the associated measure...

Page 53: ...try into the alarm list if the state of the alarm limits monitoring changes or when a breakage occurs 2mA input only Entry into the event list if the state of the prewarning limits monitoring changes the value of the individual fault currents can also be output via the analog outputs The present values of the monitored fault currents are visible via the menu of the instantaneous values Meaning of ...

Page 54: ...ture is reached PTC a short circuit or a wire sensor breakage Pt100 occurs as logic input for monitoring functions as source for digital outputs Entry into the alarm list when any state change occurs the present temperature for Pt100 measurement can also be output via analog outputs State of temperature monitoring in the instantaneous values menu PTC on the left Pt100 on the right Meaning of the u...

Page 55: ... function AND NAND OR NOR DIRECT and INVERT are available These logical functions are described in Appendix C Delay time on The time a condition must be present until it is forwarded Delay time off Time to be waited until a condition which is no longer present will be released again Description This text will be used for visualization in the alarm list List entry with data logger only Alarm event ...

Page 56: ...ed locally or via web browser or if the alarm state no longer exists By acknowledging only the flashing of the alarm symbol stops the symbol itself remains statically displayed until none of the monitoring functions is in the alarm state Logic output The summary alarm can drive an output The assignment of a digital output relay to the summary alarm is done via the settings of the corresponding out...

Page 57: ...f system events such as configuration changes power failures or reset operations and much more Events Event and alarm list Operator list Disturbance recorder Events will be registered in the disturbance recorder list By selecting the entries the course of the RMS values of all U I the wave shape of all U I during the disturbance will be recorded Events Disturbance recorder 7 7 1 Periodical data Co...

Page 58: ...be shown independently of the real averaging time Weekly profile Table Listing of all acquired mean values in the sequence of the real averaging interval The graphical representation allows comparing directly the values of the previous day or week By selecting the bars you may read the associated values Mean value Min RMS value within the interval Max RMS value within the interval Weekly display W...

Page 59: ...ed time range can be determined Selection of the meter logger group Meter content readings in table form Displaying data locally The selection works in principle in the same way as with the WEB GUI There are the following differences The individual measured quantities are arranged in a display matrix and can be selected via navigation The number of displayable meter readings is limited to 25 The t...

Page 60: ...oring functions and limit values for which state transitions need to be recorded the parameter list entry must be set to either events or alarms Displaying of event entries Event lists are a kind of logbook The occurrence of monitored events is recorded in the appropriate list with the time of its occurrence There are the following lists Alarm list Event list Operator list Example of an operator l...

Page 61: ...ded disturbances are available in the form of a logbook Each detected disturbance is entered into the disturbance recorder list with the time of its occurrence By selecting a list entry the graphical display of the measured values during this event is entered The following presentations are available Half cycle RMS curves of all voltages all currents all voltages and currents Wave shapes of all vo...

Page 62: ...ow for the selection of the quantities to display is entered by pressing OK Voltage display Current display Mixed display Display of disturbance recordings WEB GUI As with the local GUI recorded disturbances are available in the form of a logbook By selecting a list entry the graphical display of the measured values during this event is entered List of disturbance recordings Graphical display of a...

Page 63: ...iguration timeout After 5 min without interaction in a parameter selection or during entering a value in the settings menu the active configuration step is closed and the associated parameter remains unchanged The next step depends on what you have done before If the user did not change configuration parameters before the aborted step the main menu will be displayed and the device starts to monito...

Page 64: ...leaning The display and the operating keys should be cleaned in regular intervals Use a dry or slightly moist cloth for this Damage due to detergents Detergents may not only affect the clearness of the display but also can damage the device Therefore do not use detergents 8 3 Battery The device contains a battery for buffering the internal clock It cannot be changed by the user The replacement can...

Page 65: ...ed load 4 wire unbalanced load Open Y Nominal frequency 42 50 58Hz or 50 5 60 69 5Hz configurable Sampling rate 18 kHz Measurement uncertainty Reference conditions Acc IEC EN 60688 ambient 15 30 C sinusoidal input signals form factor 1 1107 no fixed frequency for sampling measurement time 200ms 10 cycles at 50Hz 12 cycles at 60Hz Voltage current 0 1 1 2 Neutral current 0 2 1 if calculated Power 0 ...

Page 66: ...Specific levels depends on the device configuration Power supply via terminals 13 14 Nominal voltage see nameplate V1 100 230V AC 50 60Hz DC 15 overvoltage category OVC III or V2 24 48V DC 15 Consumption depends on the device hardware used 27 VA 12 W Available inputs outputs and functional extensions Basic unit 1 digital input 2 digital outputs Extensions Optional modules 2 relay outputs with chan...

Page 67: ...ect measurement of a fault or earth wire current Measurement transformer Current transformer 1 1 up to 1000 1A Instrument security factor FS5 Rated output 0 2 up to 1 5 VA Measurement range IRated 1 0A max 1 2A crest factor 3 Overload 2A continuous 20A 5 x 1s interval 300s Self consumption I2 x 0 1 Ω Monitoring Alarm limit 0 03 1000 A 2 up to 100 of primary measurement range Measurement range 2mA ...

Page 68: ...ture of sensor 20 C Response delay 0 999 s Dropout delay 0 999 s Interfaces Ethernet via RJ45 socket Protocol Modbus TCP NTP http https Physics Ethernet 100BaseTX Mode 10 100 Mbit s full half duplex auto negotiation IEC61850 via RJ45 sockets 2 equivalent ports Protocol IEC61850 NTP Physics Ethernet 100BaseTX Mode 10 100 Mbit s full half duplex auto negotiation PROFINET via RJ45 sockets 2 equivalen...

Page 69: ...C Battery pack UPS 20 60 C 1 month 20 45 C 3 months 20 30 C 1 year Temperature influence 0 5 x measurement uncertainty per 10 K Long term drift 0 5 x measurement uncertainty per year Others Usage group II EN 60 688 Relative humidity 95 no condensation Altitude 2000 m max Device to be used indoor only Mechanical attributes Housing material Polycarbonate Flammability class V 0 acc UL94 non dripping ...

Page 70: ...3000V AC inputs U versus inputs I 1800V AC inputs U versus bus I O s 1 3600V AC inputs I versus bus I O s 3000V AC inputs I versus inputs I 1500V AC 1 During type test only with all protective impedances removed The device uses the principle of protective impedance for the voltage inputs to ensure protection against electric shock All circuits of the device are tested during final inspection Prior...

Page 71: ...MC Immunity for equipment used in power station and substation environment IEC EN 61131 2 Programmable controllers equipment requirements and tests digital inputs outputs 12 24V DC IEC EN 62053 31 Pulse output devices for electromechanical and electronic meters S0 output UL94 Tests for flammability of plastic materials for parts in devices and appliances 2011 65 EU RoHS EU directive on the restric...

Page 72: ...ding on the measured quantity also minimum and maximum values are determined and non volatile stored with timestamp These values may be reset by the user via display see resetting of measurements Measurement present max min 1L 2L 3Lb 3Lb P 3Lu 3Lu A 4Lb 4Lu O 4Lu Voltage U Voltage U1N Voltage U2N Voltage U3N Voltage U12 Voltage U23 Voltage U31 Zero displacement voltage UNE Current I Current I1 Cur...

Page 73: ... power factor QF3 Load factor LF Load factor LF1 Load factor LF2 Load factor LF3 cosφ H1 cosφ L1 H1 cosφ L2 H1 cosφ L3 H1 cosφ H1 quadrant I cosφ H1 quadrant II cosφ H1 quadrant III cosφ H1 quadrant IV tanφ H1 tanφ L1 H1 tanφ L2 H1 tanφ L3 H1 Umean U1N U2N 2 Umean U1N U2N U3N 3 Umean U12 U23 U31 3 Imean I1 I2 2 Imean I1 I2 I3 3 IMS Average current with sign of P Phase angle between U1 and U2 Phase...

Page 74: ... distortion component Only the fundamental reactive power may be compensated directly by means of the classical capacitive method The distortion components have to be combated using inductors or active harmonic conditioners The load factor PF is the relation between active power P and apparent power S including all possibly existing harmonic parts This factor is often called cosφ which is only par...

Page 75: ...igured for measurement in a 4 wire system with unbalanced load and the neutral connector is connected to earth In case of a single phase earth fault there is a resulting zero displacement voltage of ULL 3 The alarming may be done e g by means of a relay output Transformer secondary side Load Because in case of a fault the voltage triangle formed by the three phases does not change the voltage and ...

Page 76: ...ects occur such as additional thermal stress to operational resources or electrical mains which lead to an advanced aging or even damage Also the reliability of sensitive loads can be affected and unexplainable disturbances may occur In industrial networks the image of the harmonics gives good information about the kind of loads connected See also Increase of reactive power due to harmonic current...

Page 77: ...thod by means of the symmetrical components is ambitious and intensive to calculate The results may be used for disturbance analysis and for protection purposes in three phase systems The real existing system is divided in symmetrical system parts A positive sequence a negative sequence and for systems with neutral conductor a zero sequence system The approach is easiest to understand for rotating...

Page 78: ...y be used In case of an external synchronization the interval should be within the given range of one second up to one hour The synchronization is important for making e g the mean value of power quantities on generating and demand side comparable Trend The estimated final value trend of mean values is determined by weighted addition of measurements of the past and the present interval It serves f...

Page 79: ...onfigured meter 3 User configured meter 4 User configured meter 5 User configured meter 6 User configured meter 7 User configured meter 8 User configured meter 9 User configured meter 10 User configured meter 11 User configured meter 12 Standard meters The meters for active and reactive energy of the system are always active User configured meters To each of these meters the user can freely assign...

Page 80: ... var Reactive power phase L2 Q3 Q 3 TRMS var Reactive power phase L3 S S TRMS VA Apparent power system S1 S 1 TRMS VA Apparent power phase L1 S2 S 2 TRMS VA Apparent power phase L2 S3 S 3 TRMS VA Apparent power phase L3 F F TRMS Hz System frequency PF PF TRMS Active power factor P S PF1 PF 1 TRMS Active power factor P1 S1 PF2 PF 2 TRMS Active power factor P2 S2 PF3 PF 3 TRMS Active power factor P3...

Page 81: ... of P3 Q_MAX Q TRMS var Maximum value of Q Q1_MAX Q 1 TRMS var Maximum value of Q1 Q2_MAX Q 2 TRMS var Maximum value of Q2 Q3_MAX Q 3 TRMS var Maximum value of Q3 S_MAX S TRMS VA Maximum value of S S1_MAX S 1 TRMS VA Maximum value of S1 S2_MAX S 2 TRMS VA Maximum value of S2 S3_MAX S 3 TRMS VA Maximum value of S3 F_MM F TRMS Hz Minimum and maximum value of F UR21_MAX U neg pos UNB Maximum value of...

Page 82: ...ximum bimetal current system IB1_MAX IB 1 t3 A Maximum Bimetal current phase L1 IB2_MAX IB 2 t3 A Maximum Bimetal current phase L2 IB3_MAX IB 3 t3 A Maximum Bimetal current phase L3 Meters Name Measurement identification Unit Description ΣP_I_IV_HT P ΣHT Wh Meter P I IV high tariff ΣP_II_III_HT P ΣHT Wh Meter P II III high tariff ΣQ_I_II_HT Q ΣHT varh Meter Q I II high tariff ΣQ_III_IV _HT Q ΣHT v...

Page 83: ...ata available only if the device is equipped with at least one optional fault current module ϑ m 1 m 2 Graphic Present measurements and states of temperature monitoring Data available only if the device is equipped with at least one temperature module MT_P_I_IV Graphic mean value P I IV Trend last 5 interval values minimum and maximum MT_P_II_III as MT_P_I_IV Graphic mean value P II III Trend last...

Page 84: ... 50th Total Harmonic Distortion of all voltages HO_UX_MAX as HO_IX Graphic Maximum values odd harmonics 3rd up to 49th Total Harmonic Distortion of all voltages HO_IX_MAX as HO_IX Graphic Maximum values odd harmonics 3rd up to 49th Total Harmonic Distortion of all currents HE_UX_MAX as HO_IX Graphic Maximum values even harmonics 2nd up to 50th Total Harmonic Distortion of all voltages HE_IX_MAX as...

Page 85: ...2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR_M1 M1_MM M2 TR_M2 M2_MM M3 TR_M3 M3_MM M4 TR_M4 M4_MM M5 TR_M5 M5_MM M6 TR_M6 M6_MM M7 TR_M7 M7_MM M8 TR_M8 M8_MM M9 TR_M9 M9_MM M10 TR_M10 ...

Page 86: ...2_TRIANGLE PF_MIN Cφ_MIN I 1 1 1 2 I 2 1 2 2 ϑ 1 1 1 2 ϑ 2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR_M1 M1_MM M2 TR_M2 M2_MM M3 TR_M3 M3_MM M4 TR_M4 M4_MM M5 TR_M5 M5_MM M6 TR_M6 M6_MM...

Page 87: ...2 I 2 1 2 2 ϑ 1 1 1 2 ϑ 2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR_M1 M1_MM M2 TR_M2 M2_MM M3 TR_M3 M3_MM M4 TR_M4 M4_MM M5 TR_M5 M5_MM M6 TR_M6 M6_MM M7 TR_M7 M7_MM M8 TR_M8 M8_MM M9...

Page 88: ...1 1 2 ϑ 2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR_M1 M1_MM M2 TR_M2 M2_MM M3 TR_M3 M3_MM M4 TR_M4 M4_MM M5 TR_M5 M5_MM M6 TR_M6 M6_MM M7 TR_M7 M7_MM M8 TR_M8 M8_MM M9 TR_M9 M9_MM M10...

Page 89: ...Cφ_MIN I 1 1 1 2 I 2 1 2 2 ϑ 1 1 1 2 ϑ 2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR_M1 M1_MM M2 TR_M2 M2_MM M3 TR_M3 M3_MM M4 TR_M4 M4_MM M5 TR_M5 M5_MM M6 TR_M6 M6_MM M7 TR_M7 M7_MM M8...

Page 90: ...1 2 I 2 1 2 2 ϑ 1 1 1 2 ϑ 2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR_M1 M1_MM M2 TR_M2 M2_MM M3 TR_M3 M3_MM M4 TR_M4 M4_MM M5 TR_M5 M5_MM M6 TR_M6 M6_MM M7 TR_M7 M7_MM M8 TR_M8 M8_MM ...

Page 91: ...2 ϑ 2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR_M1 M1_MM M2 TR_M2 M2_MM M3 TR_M3 M3_MM M4 TR_M4 M4_MM M5 TR_M5 M5_MM M6 TR_M6 M6_MM M7 TR_M7 M7_MM M8 TR_M8 M8_MM M9 TR_M9 M9_MM M10 TR_...

Page 92: ...P3_MAX Q3_MAX S3_MAX PF P Q S P_MAX Q_MAX S_MAX P_TRIANGLE P1_TRIANGLE P2_TRIANGLE P3_TRIANGLE PF_MIN Cφ_MIN I 1 1 1 2 I 2 1 2 2 ϑ 1 1 1 2 ϑ 2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR...

Page 93: ..._MAX S3_MAX PF P Q S P_MAX Q_MAX S_MAX P_TRIANGLE P1_TRIANGLE P2_TRIANGLE P3_TRIANGLE PF_MIN Cφ_MIN I 1 1 1 2 I 2 1 2 2 ϑ 1 1 1 2 ϑ 2 1 2 2 ΣP_I_IV_HT ΣP_I_IV_NT ΣP_II_III _NT ΣP_II_III_HT ΣQ_I_II_HT ΣQ_I_II _NT ΣQ_III_IV _HT ΣQ_I_II _NT ΣMETER1 ΣMETER2 ΣMETER3 ΣMETER4 ΣMETER5 ΣMETER6 ΣMETER7 ΣMETER8 ΣMETER9 ΣMETER10 ΣMETER11 ΣMETER12 MT_P_I_IV MT_P_II_III MT_Q_I_II MT_Q_III_IV MT_S M1 TR_M1 M1_MM...

Page 94: ...ions is not fulfilled OR A B Y 0 0 0 0 1 1 1 0 1 1 1 1 Function is true if at least one of the input conditions is fulfilled NOR A B Y 0 0 1 0 1 0 1 0 0 1 1 0 Function is true if none of the input conditions is fulfilled Using DIRECT or INVERT the input is directly connected to the output of a monitoring function without need for a logical combination For these functions only one input is used DIR...

Page 95: ...mmunications However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving antenna Increase t...

Page 96: ...c components AND 94 DIRECT 94 INVERT 94 NAND 94 NOR 94 OR 94 Logic functions 94 M Measured quantities 72 Basic measurements 72 Bimetal current 78 earth fault monitoring 75 harmonic analysis 76 Load factors 74 mean values and trend 78 meters 79 system imbalance 77 zero displacement voltage 75 Measurement displays 46 Measurements reset 48 Mechanical mounting 7 Menu operation 45 N NTP 35 O Operating ...

Reviews: