55/66
C4 Operation
C4.1 Measuring current
Read safety section of instructions before operating this product.
Ensure conductor to be tested is de-energised
Connect the output of the electronics to the input of an oscilloscope or other data recording
device.
The flexible current probe is not for use on conductors with a potential of over 1000V.
Wrap the flexible probe around the conductor to be tested, close coupling. Energise the circuit
under test. Locate the coupling away from nearby conductors.
Always use appropriate equipment for personal protection. Do not use the flexible
current probe to measure bare conductors unless wearing protective clothing
suitable for high voltage work.
C4.2 Operation
To activate unit, move the rotary switch from the “off” position to the required measuring range.
If the value of current being measured is unknown, first select the highest range and then
reduce accordingly.
C5 Maintenance
Do not use ACP 3000_5 or ACP 3003_5 if damaged.
Always inspect the integrator unit, connecting cable, and flexible probe for damage before use.
To avoid electric shock, keep the ACP clean and free of surface contamination.
Use Isopropyl alcohol to clean the electronics unit and the probe
Make sure the flexible probe, connecting cable, and electronics enclosure are dry before further
use