Install & Operating Manual
CamSafe
11
Option : Mobile Integrity Probe (MIP)
The Mobile Integrity Probe (MIP) is a manual scanning system for for HEPA filter validation purposes,
fully integrated into the housing. It consists of a static upstream concentration probe and -port as well
as a mobile downstream sampling probe and -port.
➔
A comprehensive testing report on system performance is available, upon request.
➔
For scanning speed and sampling rates with respect to the airflow, please refer to ISO 14644-3
or other applicable standard.
•
For reliable results and an ISO 14644-3 compliant aerosol distribution, we recommend using the
Camfil Aerosol Injection Module,
see § 12
PLEASE NOTE:
•
All generally applicable, statutory and other regulations relating to environmental protection
and accident prevention must be observed.
•
All safety regulations applicable to the respective area of use must be observed.
This includes protective clothing (gloves, protective suits, protective goggles, breathing
protection).
•
CamSafe safety housings with particle measuring system must be operated
on the extraction side only.
•
The particle measuring system may only be operated by qualified specialist staff.
The operator is responsible for the necessary training and instruction of the specialist staff
commissioned by him/her.
•
The operator must ensure that the CamSafe particle measuring system is only operated
in proper condition.