63
Scanner
®
2000 microEFM
Section 3
Power Supply Wiring
Internal Power Supply
The Scanner 2000 microEFM is shipped with a lithium battery pack. To supply power to the instrument,
connect the battery cable to connector J1 on the main circuit assembly (Figure 3.2).
Low-power microprocessor technology enables the Scanner 2000 to operate for an estimated 1 year on a
lithium battery pack. The lithium battery pack is ideal for use in extreme temperatures, although extreme cold
temperatures may reduce battery life.
To maximize battery life,
• operate the Scanner using the following default configuration settings:
–
calculation frequency: 1 minute
–
logging frequency (interval): 1 hour
–
download frequency: monthly
• disconnect the Scanner 2000 from the RS-232 to RS-485 converter when ModWorX Pro software is not
in use. When ModWorX Pro is running, the computer powers the converter; when the software is not run
-
ning, the Scanner 2000 powers the converter, causing a current drain to the Scanner battery.
• avoid the following conditions/activities:
– operation at extremely cold temperatures
–
use of digital output (pulse or alarm)
–
use of analog input without external power (when expansion board is installed)
J1
T
B1
TB2
TB3
BATTERY
13
14
J2
P
O
W
E
R
R
TD
P
O
R
T 2
P
O
R
T 1
I
+
R
+
R
-
I
-
SCANNER 2000
Main Circuit Board
PN: 9A-30160010
–
+
–
+
–
+
–
+
2
3
4
5
6
1
7
8
9
10
11
12
TF
M
1
SWITCH
DIG OUT 1
Figure 3.2—Lithium battery pack connection
Summary of Contents for Scanner 2000 microEFM
Page 1: ...NUFLO Scanner 2000 microEFM Hardware User Manual Manual No 9A 30165023 Rev 10 ...
Page 26: ...26 Section 1 Scanner 2000 microEFM ...
Page 104: ...A 18 Appendix A Scanner 2000 microEFM ...
Page 106: ...B 2 Appendix B Scanner 2000 microEFM ...