A-3
MC-III™ EXP Flow Analyzer
Appendix A
Additionally, when the log viewing function times out, the accumulated total is automatically saved to
memory.
Saving the Total.
When the daily log display times out, the MC-III EXP will automatically save the
accumulated total.
Explosion-Proof Communications Adapter
The explosion-proof communications adapter provides an RS-485 connection for connecting a laptop or PC to
the instrument without removing the instrument cover. When the adapter is ordered with an MC-III EXP, it is
factory installed. The installed adapter is comprised of an RS-485 adapter socket, a blanking plug, and a union
nut (Figure A.4). It may be relocated to either conduit opening in the instrument housing.
Caution:
This adapter is available in two models—a plated steel CSA-approved model and a brass
ATEX-approved model.
An RS-232 to RS-485 converter cable (available from Cameron’s Measurement Systems Division) is required
for connecting the adapter to a laptop computer. See
MC-III Parts List (CSA), page 97,
and
for converter cable options.
Adapter socket
Union nut
Blanking plug
Figure A.4—Components of explosion-proof communications adapter
Assembly of the Plug Connector
A plug connector that mates with the RS-485 adapter socket is shipped with the device (uninstalled). This
plug connector should be wired to an RS-485 converter cable as shown in
and stored
with the cable when the COM adapter is not in use.
Summary of Contents for NUFLO MC-III EXP
Page 1: ...NUFLO TM MC III EXP Flow Analyzer User Manual Manual No 9A 50165003 Rev 06 Valves Measurement ...
Page 18: ...12 Section 1 MC III EXP Flow Analyzer ...
Page 35: ...29 MC III EXP Flow Analyzer Section 2 ...
Page 46: ...40 Section 3 MC III EXP Flow Analyzer ...
Page 81: ...75 MC III EXP Flow Analyzer Section 4 ...
Page 93: ...87 MC III EXP Flow Analyzer Section 5 Figure 5 17 Dropdown menu for exporting log data ...
Page 94: ...88 Section 5 MC III EXP Flow Analyzer ...
Page 105: ...99 MC III EXP Flow Analyzer Section 6 ...
Page 114: ...A 9 MC III EXP Flow Analyzer Appendix A ...
Page 121: ...B 7 MC III EXP Flow Analyzer Appendix B ...
Page 123: ...C 2 Appendix C MC III EXP Flow Analyzer ...
Page 141: ...D 18 Appendix D MC III EXP Flow Analyzer ...
Page 143: ......