Cambridge Sensotec Rapidox 1100E Oxygen Analyser Instruction Manual
Version 3.8
38
Last printed 29/08/2012 16:01:00
that your COM ports are recognised by your computer. Check in your
Windows device manager to see if there are any conflicts. If you have an
internal modem fixed to your PC you may experience difficulties using this
software. Contact Cambridge Sensotec for further advice.
Q:
I selected ‘Write to the Analyser’ and now the Rapidox is way off
calibration.
A:
Only select ‘Write to the Analyser’ once all the values are correctly entered
in the boxes on the right. The best procedure is to ‘Read Analyser
Configuration’ and then press ‘Copy’ so the values in the boxes are the same
as those stored in the Rapidox memory. You can then edit these values and
then select ‘Write to the Analyser’ when finished.
Q:
How do I get back to the factory calibration for my instrument.
A:
Select ‘Default Configuration’ from the ‘File’ menu and load the default ‘rxc’
file from the Rapidox software programme folder on your PC, or look for the
default ‘rxc’ file on your CD. Select this file and then select ‘Write to the
Analyser’. This will load the factory calibration back into the Rapidox.
Alternatively you can load a standard default configuration any time using
the ’10-Load Defaults’ option from the user menu which is accessed via the
front panel keypad. This configuration will not be perfect for your analyser
but will be good enough to get you started again.
10
Warranty
The Rapidox analyser has been carefully tested and inspected before shipment
and is guaranteed to be free from defective materials and workmanship for a
period of twelve months from date of purchase. The sensor head is replaceable
and has a life expectancy of 3-5 years. However, if the analysis gas contains
corrosive gases or large quantities of particulates, sensor life may be shortened. In
the case of the latter, it is normally possible to insert a filter prior to the sensor
head to remove the particulate material.
10.1
Conditions of Warranty:
1) This warranty is in addition to and does not affect any statutory rights of
consumer purchasers. This warranty is valid worldwide on a “back to base basis”.
2) This warranty covers breakdowns due to design or manufacturing faults; it does
not apply to damage, however caused, wear and tear, neglect, unauthorised
adjustment or repair, or any items of limited natural life.
3) In the event of failure, please take the following action:
a] Refer to the “Troubleshooting” section of your instruction manual to
identify and possibly correct the problem.
Summary of Contents for SENSOTEC Rapidox 1100E
Page 46: ......