Cambridge Sensotec Rapidox 1100E Oxygen Analyser Instruction Manual
Version 3.8
19
Last printed 29/08/2012 16:01:00
To read the current configuration stored in your Rapidox select ‘Read Analyser
Configuration’ from the ‘File’ menu, or click the ‘Read’ button on any of the
‘Config’ pages. Note that the entries in the boxes on the RHS (under the ‘Write’
button) will turn red if the value is different to that just read and a red asterisk
will appear next to the ‘Write’ button to warn you that a change has been made.
The left-hand set of grey text fields (‘read fields’) will be updated with the current
configuration data. To save this information (e.g., if several people share the same
instrument) click the long copy (‘>>’) button (between the ‘Read’ and ‘Write’
columns) on any of the ‘Config’ pages in order to transfer the data into the edit
boxes, select ‘Save Configuration as’ from the ‘File’ menu and choose a filename.
This data set can be subsequently reloaded at any time by selecting ‘Load
Configuration File’ from the ‘File’ menu and then selecting the folder where the
configuration files have been stored.
If at any time you wish to restore the Rapidox to its factory default settings select
‘Default Configuration’ from the ‘File’ menu to load these values. The software
will search for a unique ‘.rxc’ file which is loaded onto the PC during the software
installation. The filename is ‘XXXXXXX.rxc’ where ‘XXXXXXX’ is the unit’s
serial number, which can be found on the rear of the machine. The original copy
of this file is also on the CD-ROM that came with the analyser. Normally this file
is automatically copied to the main Rapidox folder, but if the software cannot
locate it then it will ask you for the serial number to help it search. The ‘rxc’ file
will be loaded into the ‘Write’ columns, and the software will then automatically
perform a ‘Read’ to update the data in the ‘Read’ columns in order to highlight the
differences between the current configuration (in the ‘Read’ columns) and the
settings from the file (in the ‘Write’ columns). If you wish to proceed with the
reconfiguration simply click any of the three ‘Write’ buttons to reprogramme the
Rapidox.
The right-hand set of white editable text fields (‘write fields’) is used to enter new
values to be programmed into the Rapidox. The values in these fields can either
be entered manually, or copied across from the ‘read fields’ by clicking the vertical
‘>>’ (copy) button, and then edited as required, however greyed boxes under the
‘Write’ buttons are not editable as they contain calibration data; they can only be
changed by copying data across from the ‘Read’ fields. Note that when you click
on the copy button all three ‘Config’ pages are updated at the same time and there
is no need to repeat the action when you toggle to another page. Alternatively,
you can load saved configuration data by selecting ‘Load Configuration File’ from
the ‘File’ menu and selecting the required file. In order to programme the analyser
select ‘Write configuration to analyser’ from the ‘File’ menu, or click the ‘Write’
button on any ‘Config’ page.
7.5
Reconfiguring the Analyser
The on-screen edit boxes contain variables that can be reprogrammed into the
Rapidox in exactly the same manner as using the front Keypad. For a full
description please refer to the relevant section in this manual.
Summary of Contents for SENSOTEC Rapidox 1100E
Page 46: ......