AFX SERIES® OPERATION MANUAL
SECTION 8:
Entire Contents Copyright
2018 by Pacific Power Source, Inc. (PPS)
• All Rights Reserved • No reproduction without written authorization from PPS.
AFX Series Power Source Operation Manual
Page 253 of 402
Command Syntax
[SOURce:]IPROTect:STATe
Description
This command turns the current protection mode on or off
Parameters
[ 0 | OFF | 1 | ON ]
Parameter Format
<b>
Example
IPROT:STAT ON
Query Format
[SOURce:]IPROTect:STATe?
Returned Data Format
<b>
Query Example
IPROT:STAT?
1
Query Format
[SOURce:]IPROTect:TRIPped?
Description
This command returns the current protection state. If the peak
current protection has been tripped, a 1 is returned.
Returned Data Format
<b>
Query Example
IPROT:TRIP?
0
Command Syntax
[SOURce:]IPEAK:LIMit
Description
This command sets the peak current trip level.
Parameters
0.0000 – 100000.0000
Parameter Format
<nr2>
Example
IPEAK:LIM 72.25
Query Format
[SOURce:]IPEAK:LIMit?
Returned Data Format
<nr2>
Query Example
IPEAK:LIM?
72.2500
8.7.6
Phase Programming Commands
Command Syntax
[SOURce:]PHASe#
Description
This command programs the phase angle for the selected phase. A
phase reference (#) must be specified.
Note that unless external sync mode is used, the A phase (# = 1)
cannot be programmed is fixed at 0 degrees.
Parameters
0.0 – 359.9
Parameter Format
<nr2>
Example
PHAS2 122.5
Query Format
[SOURce:]PHASe#?
Returned Data Format
<nr2>
Query Example
PHAS2?
122.5