User Manual - Rev AJ
California Instruments
9.4.6.1 DIPS AND INTERRUPTIONS TEST
Scroll to the DIPS AND INTERRUPTIONS entry using the up and down cursor keys. Press the
ENTER key to select the DIPS AND INTERRUPTIONS menu. The screen will appear as shown
in Figure
Figure 9-15: IEC Dips and Interrupts
This field enables or disables the -411 test mode.
ATE
This field is not available on MX Series systems.
L
The RUN ALL selection will cause the following automated test sequence sugg
STATE
EOS ST
RUN AL
ested by the
standard to be run:
Step
Output in
% of U
T
No of Cycles
Start angle
(Degrees)
Repeat #
times
Delay
between
repeats (s)
1 0 0.5
0
3
10
2 0 0.5
180
3
10
3
0
1
0,45,90
3 at diff ø
10
4
0
5
45,90,135
3 at diff ø
10
5
0
10
90,135,180
3 at diff ø
10
6
0
25
180,225,270
3 at diff ø
10
7
0
50
270,315,0
3 at diff ø
10
8 40 0.5
0
3
10
9 40 0.5
180
3
10
10
40
1
0,45,90
3 at diff ø
10
11
40
5
45,90,135
3 at diff ø
10
12
40
10
90,135,180
3 at diff ø
10
13
40
25
180,225,270
3 at diff ø
10
14
40
50
270,315,0
3 at diff ø
10
15 70
3
10
0.5
0
16 70
0.5
180
3
10
17
70
1
0,45,90
3 at diff ø
10
18
70
5
45,90,135
3 at diff ø
10
19
70
10
90,135,180
3 at diff ø
10
20
70
25
180,225,270
3 at diff ø
10
21
70
50
270,315,0
3 at diff ø
10
Table 9-19: Dips and Interruptions Tests Performed During RUN ALL
The user can change the NOMINAL Ut voltage for this. The RUN ALL Command line will change
to ABORT during the test. Selecting ABORT and pressing the ENTER key will terminate the test
at any time and the output voltage will return to the nominal value.