User and Programming Manual - Rev N
10.9.7 *SAV
This command stores the present state of the AC source to a specified location in memory.
The RP Series offers eight non-volatile memory locations for storing instrument setups.
Command Syntax
*SAV
Parameters
0 through 7
Related Commands
*RCL *RST
10.9.8 *SRE
This command sets the condition of the Service Request Enable Register. This register
determines which bits from the Status Byte Register (see *STB for its bit configuration) are
allowed to set the Master Status Summary (MSS). A 1 in any Service Request Enable
Register bit position enables the corresponding Status Byte Register bit and all such
enabled bits then are logically OR-ed to cause Bit 6 of the Status Byte Register to be set.
See Section 12 for more details concerning this process.
When the controller conducts a STB? command, the MSS bit is not cleared. When *SRE is
cleared (by programming it with 0), the source cannot generate an MSS bit.
Command Syntax
*SRE <NRf>
Parameters
0 to 255
Default Value
128
Example *SRE
255
Query Syntax
*SRE?
Returned Parameters
<NR1>(Register binary value)
Related Commands
*ESE *ESR
2001RP April
2007
107
Summary of Contents for 2001RP
Page 33: ...User and Programming Manual Rev N Figure 3 7 Test Setup 2001RP 24 April 2007 2001RP...
Page 47: ......
Page 60: ...User and Programming Manual Rev N Figure 6 3 Internal Adjustments 2001RP April 2007 51...
Page 65: ...User and Programming Manual Rev N Figure 7 1 Assembly Location 56 April 2007 2001RP...
Page 68: ...User and Programming Manual Rev N Figure 7 2 Replaceable Fuse Locations 2001RP April 2007 59...
Page 70: ...User and Programming Manual Rev N Figure 8 1 Rear Panel View 2001RP April 2007 61...
Page 90: ......