2
:
Re-Energy
3
:
AHour
4
:
WHour
5
:
Reset Meas
6
:
Reset ReE
•
UUT Temp: temperature measurement value.
•
Re-Energy: the regenerative power information.
•
AHour: the capacity information of the present discharge or charge
test, which is also cleared when the battery test mode or battery
simulation mode is turned on.
•
WHour: the power statistics information, when the battery test mode
or battery simulation mode is turned on, the data will also be cleared.
•
Reset Meas: The zero-clearing operation, when the instrument is re-
powered or the source /load is switched it will be also performed.
•
Reset ReE: The zero-clearing regenerative power information
operation.
d. Press
[Esc]
to exit.
The information will display on the home screen when the Ext-Meter Disp
set to On in system menu, the operation refers to
Temperature and Total Historical Regenerative Power
.
12.000V
10.000A
38.0
℃
1000.0Wh
7.6 Observe the Regenerative power and
Capaticy
User can observe the total historical regenerative power and capacity
information on screen.
1. Press the composite keys
[Shift]+[P-set]
(System) to enter the system
menu.
2. Rotate the knob to select the
External Meter
and press
[Enter]
to confirm.
3. The DUT temperature, total historical regenerative power and capacity are
displayed on screen.
1
:
UUT Temp
2
:
Re-Energy
3
:
AHour
4
:
WHour
5
:
Reset Meas
6
:
Reset ReE
•
UUT Temp: indicates the DUT temperature.
Copyright © Itech Electronic Co., Ltd.