Section 12
Model 601/601P/602/602P/603 NDIR Analyzer
California Analytical Model 600_NDIR_Series_v10.0 CE
Operators Manual
Page 82 of 86
12.5.9.
SARE:
Auto range on
Command Response
Description
_SARE_K0
_SARE_s
Set auto range on for all channels
_SARE_Km
_SARE_s
Set auto range on for channel m
12.5.10.
SARA:
Auto range off
Command Response
Description
_SARA_K0
_SARA_s
Set auto range off for all channels
_SARA_Km
_SARE_s
Set auto range off
12.5.11.
SREM:
Remote mode for AK-commands
Command Response
Description
_SREM_K0
_SREM_s
Set device in remote mode
12.5.12.
SMAN:
Manual control to control device manually
Command Response
Description
_SMAN_K0
_SMAN_s
Set device in manual mode
12.5.13.
SMGA:
Start measuring
Command Response
Description
_SMGA_K0 _SMGA_s
Start
measuring
Open all sample valves
_SMGA_Km
_SMGA_s
Open sample valve of channel m
12.5.14.
SNKA:
Saves measured value as new offset.
Command Response
Description
_SNKA_K0
_SNKA_s
Saves measured value of actual range
for each channel as new offset if zero
valve is opened
_SNKA_Km
_SNKA_s
Saves measured value of actual range
as new offset if zero valve is opened
12.5.15.
SEKA:
Saves measured value as new span value
Command Response
Description
_SEKA_K0
_SEKA_s
Saves new span values for each
channel if span valve is opened
_SEKA_Km
_SEKA_s
Saves measured value of actual range
as new span value if span valve is
opened
12.5.16.
SUDP:
Start /Stop UDP data streaming
Command Response
Description
_SUDP_K0_ON _SUDP_s
Start
Data streaming via the UDP
channel. You need to configure the
channel before with EUDP command
_SUDP_K0_OFF _SUDP_s
Stop
streaming via the UDP channel
Summary of Contents for 600-NDIR
Page 1: ......