background image

CAEN

 

 

Electronic Instrumentation

 

 

A7030-AG7030 3kV/ 1mA (1.5W) HV Boards 

Channel Characteristic Table 

Table 1 

 Channel characteristics of the Mod. A7030 / AG7030 HV Board 

Polarity 

Positive / Negative depending on purchased version  

Output Voltage 

0 ÷ 3 kV 

Max. Output Current 

1mA 

VSet Resolution 

50 mV  

VMon Resolution 

10 mV 

Current Set Resolution 

20 nA 

Current Monitor Resolution 

2 nA 

IMAX hardware 

0÷1 mA 

IMAX hardware resolution 

1 µA 

IMAX hardware accuracy 

<

 1% of FSR 

VMAX hardware 

0 ÷ 3 kV common for all the board channels 

VMAX hardware resolution 

1 V 

VMAX hardware accuracy 

<

 1% of FSR 

VMAX software 

0 ÷ 3 kV settable for each channel 

VMAX software resolution 

1 V 

Ramp Up / Ramp Down 

1÷ 500 Volt/sec, 1 Volt/sec step 

Voltage Ripple 

10 ÷ 1000 Hz 

<15 mV typical;  20 mV max  

 

1000 Hz 

<5 mV typical;  10 mV max 

VMon vs. VOut Accuracy 

3,4

 

typical: ± 0.3% ± 0.2 V  

max: ± 0.3% ± 1 V 

VSet vs. VMon Accuracy 

2

 

typical: ± 0.3% ± 0.2 V  

max: ± 0.3% ± 1 V 

IMon vs. IOut Accuracy 

2

 

typical: ± 1% ± 100 nA  

max: ± 1% ± 1 µA 

Maximum output power 

1.5 W (per channel)  

Consumption @ full power 

48 channels (A7030): 160 W  

36 channels (A7030T): 120 W  

24 channels (A7030S, A7030L): 90 W  

12 channels (A7030D): 50 W  

 

 
 

2

 From 10 Hz to 20 MHz at full load; ripple may exceed such limits whenever OVC and UNV occur (see Output control and 

monitoring) 

3

 From 10% to 90% of Full Scale Range 

4

 During operation in Overcurrent or when VMAX Hardware is reached (and/or exceeded), VMON values have to be assumed as 

“indication”; possible monitor drifts are caused by the different regulation mode.

 

 

Summary of Contents for A7030

Page 1: ...CAEN Tools for Discovery n Electronic Instrumentation Rev 5 Es A7030 AG7030 3kV 1mA 1 5W HV Boards Rev 7 03 September 2019...

Page 2: ...t is defective due to workmanship or materials and has not been caused by mishandling negligence on behalf of the User accident or any abnormal conditions or operations CAEN declines all responsibilit...

Page 3: ...cautions 9 Safety Terms and Symbols on the Product 9 Installation 9 3 Operating modes 10 Output control and monitoring 10 Channel STATUS Flag 10 Interlock protection 11 Grounding specifications 11 Saf...

Page 4: ...red via software if exceeded the relevant channel can be programmed either to turn off after a programmable trip time or to keep on providing the maximum allowed current this particular feature allows...

Page 5: ...esolution 1 V Ramp Up Ramp Down 1 500 Volt sec 1 Volt sec step Voltage Ripple 2 10 1000 Hz 15 mV typical 20 mV max 1000 Hz 5 mV typical 10 mV max VMon vs VOut Accuracy 3 4 typical 0 3 0 2 V max 0 3 1...

Page 6: ...CAEN Electronic Instrumentation A7030 AG7030 3kV 1mA 1 5W HV Boards 6 Front Panel Fig 1 A7030 family front panel 48 36 24 Multipin 24 SHV and 12 SHV channel version...

Page 7: ...ge VMAX common to all the channels Its value can be read out via software IMAX Function it allows to adjust the hardware maximum current IMAX common to all the channels Its value can be read out via s...

Page 8: ...CH26 37 CH23 48 CH25 7 CH32 17 CH29 28 CH31 38 CH28 49 CH30 8 CH37 18 CH34 29 CH36 39 CH33 50 CH35 9 CH42 19 CH39 30 CH41 40 CH38 51 CH40 10 CH47 20 CH44 31 CH46 41 CH43 52 CH45 21 RETURN 42 SAFETY L...

Page 9: ...rs To avoid electric shock or fire hazard do not operate this product with covers or panels removed Do Not Operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or...

Page 10: ...POn EN DIS Power On option which can be enabled or disabled If this option is enabled at Power On or after a Restart each channel is restored in the same condition defined by the Power parameter it wa...

Page 11: ...pin pin 21 on A AG7030 7030T pin 10 on A AG7030L if the board features SHV connectors skip this step Then it is necessary to configure the Interlock connector as follows contact open INTERLOCK voltage...

Page 12: ...not a working one The Connector Configurator allows to optimize the connection of the return and of DGND Earth The best configuration must be determined by the user upon application the optimal conne...

Page 13: ...ustomers including engineers scientists and technical professionals who all trust them to help achieve their goals faster and more effectively CAEN S p A CAEN GmbH CAEN Technologies Inc Via Vetraia 11...

Reviews: