Page 5-19
NAVIGAT X MK 1 Mod. 10 / SR-180 MK 1 Mod. 10
056310
3-149 MK110ENG.P65
19) Insert the syringe with the cannula into
the recess of the gyrosphere. Draw up the
mercury and insulating fluid completely.
20) Inject the mercury and insulating fluid into
the empty phial.
21) Placing a thumb in the recess of the
gyrosphere, lift the gyrosphere out of the
container and pour the remaining support-
ing fluid from the recess into the plastic
bowl.
22) Place the gyrosphere back into the lower
section of the gyrosphere container.
23) If there is any mercury left in the plastic
bowl, take it up with the syringe and inject
it into the phial.
24) Close the phial and put it back into the
box.
25) Return the box with the phial to C. PLATH
for recycling.
Note:
Used mercury and insulating fluid may not
be reused. When replacing the gyro-
sphere, always use fresh mercury and in-
sulating fluid.