background image

GasAlertMicroClip 

Quick Reference Guide 

Alarms 

Alarm 

Low Alarm 

Slow siren 
Slow alternating flash 

llmlDI  and gas bar flash 
Vibrator alarm activates 

High Alarm 

Fast siren 
Fast alternating flash 

llmlDI  and gas bar flash 
Vibrator alarm activates 

Multi-Gas Alarm 

Alternating low and high alarm siren and 

flash 
llmlDI  and gas bars flash 
Vibrator alarm activates 

Display 

-'fi.1D9D' 

17 

,20�-

,i�-

If Stealth mode is enabled, the audible and  visual  alarms  are 

disabled. Only the  vibrator  alarm activates .. 

TWAAlarm 

Slow siren 

Alann 

Slow alternating flash 

llmlDI and gas bar flash 
Vibrator alarm activates 

STELAlarm 

Fast siren 
Fast alternating flash 

llmlDI  and gas bar flash 
Vibrator alarm activates 

Over Limit (OL) Alarm 

Fast siren and alternating flash 

llmlDI and gas bar flash 
Vibrator alarm activates 

OL displays 

Display 

-'fi.1D9D' 

D  D 

20�- □t: 

Alarm 

Sensor Alarm 

During startup Error [sensor name] 

displays 

During  normal operation Err displays 

Low B attery Alarm 

Sequence of 10 rapid sirens and 

alternating flashes with 7 seconds of 

silence in between (continues for 

15 minutes) 
=::JI and !lmlDI flash, LOW BAT 

displays, and the vibrator alarm activate, 

After 15 minutes of the Low Battery 

alarm, the Automatic Shutdown Alarm 

sequence begins 
OFF displays before deactivating 

Display 

(,-,-

.. Q 

20�-

IJ

L,  'fi.1D9D' 

I� 

I

J':: D  '\ Di 

120��  .. Cl 

Alann 

Confidence B eep and lntelliFlash 

One beep and flash every second 

Note: Confidence Beep and 

lntelliFlash automatically 

deactivate during low battery 

alann, self-test fail, calibration fail, 

bump test fail, 

or 

an alann event. 

Automatic Shutdown Alarm 

Sequence of 1 O rapid sirens and 

alternating flashes with 1 second of 

silence in between (sequence reactivates 

seven times) 
LOW BAT and l!!!.'!il!ll display, and 

vibrator alarm activates 

OFF displays before deactivating 

GasAlertMicroClip 

Alarms 

Display 

20�-

oe: 

'fi.1D9D' 

LOW 

BRT, 

'Note: Alarms can be set to latching or non-latching. To enable/disable 

atching, select/deselect 

Latching Alarms 

in Fleet Manager II. Local 

regulations may require 

latching alarm. 

/flow Alarm Acknowledge 

is 

enabled and 

low alarm occurs, press 

'.) lo 

disable the audible alarm. The visual and vibrator alarms remain 

activated. 

If 

the alarm escalates 

lo a 

high, STEL, 

or 

1WA 

alarm, the 

audible  alarm reactivates. 

GasA/ertMicroClip 

Quick Reference Guide 

User Options Menu 

To modify the user options, connect the detector to the IR Link 

adapter and open Fleet Manager II. Refer to the Fleet Manager 

II Operator's Manual for complete instructions. 

1.  Sensors (H2S, CO, LEL, and 02) 

Sensor Disabled: Disables the sensor. 

Calibration Gas (ppm) / (%LEL) / (%02): Defines the 

calibration gas concentration for each sensor. 

Calibration Interval (days): Defines how often a 

calibration should be perfonned. 

Bump Interval (days): Defines how often a bump test 

should be perfonned. 

Low Alarm (ppm) / (%LEL) / (%02): Defines the low 

alann setpoint. 

High Alarm (ppm) / (%LEL) / (%02): Defines the high 

alann setpoint. 

TWA Alarm (ppm): Defines the time-weighted average 

(TWA) alann setpoint. H2S and CO sensors only. 

STEL Alarm (ppm): Defines the short-tenn exposure limit 

(STEL) alarm setpoint. H2S and CO sensors only. 

STEL Interval (minutes): Defines the short-term 

exposure limit (5-15 minutes). H2S and CO sensors only. 

Auto-Zero Startup: When enabled, the detector 

automatically zeros the H2S, CO, and LEL sensors during 

the startup self-test. 

LEL By Volume CH4: When enabled, the LEL reading is 

displayed as %vol.,  assuming a methane  environment. 
02 Auto-Calibration on Startup: When enabled, the 02 

sensor is automatically calibrated during startup. 

2.  User Options 

Confidence Beep: When enabled, the detector beeps 

once every second to verify the battery has sufficient 

power to detect hazardous gas and emit an alarm. 

Detector is shipped with Confidence Beep disabled. 

Latching Alarms: When enabled, the audible, visual, and 

vibrator alarms persist during a high or low alarm until the 

gas concentration is below 

the 

low alarm setpoint and has 

been acknowledged by pressing 

0-

Safe Mode: When enabled, SAFE continuously displays 

on the LCD when all gas concentrations are normal or 

below the alarm setpoints. 

Stealth Mode: When enabled, the audible alarm, LEDs 

and backlight are disabled. :tic displays on the LCD. 

Low Alarm Acknowledge: When enabled, the audible 

alarm can be disabled during a low alarm. T he vibrator, 

LEDs, and LCD remain enabled. For H2S, CO, and LEL 

sensors only. 

Datalog Interval (seconds): Enter a value 

(1-120 seconds). 

Force Calibration When Overdue: When enabled, the 

detector automatically enters calibration during startup for 

overdue sensors. If the sensors are not calibrated 

immediately, the detector will deactivate. 

Cal Lock: W�en enabled, the sensors can only be 

calibrated  using an IR device (IR Link with Fleet Manager 

II 

or the MicroDock 

II 

base station). 

Force �ump When Overdue: When enabled, a bump test 

1s required 1f the sensor has exceeded its bump test 

interval. If a successful bump test is not performed  the 

detector will deactivate. 

lntel_liFlash:_ When e_nabled, the green LED flashes to 

provide continuous visual confirmation that the detector is 

operating correctly. 

lntelliFlash  automatically deactivates during a low 

battery alarm, a self-test fail, a calibration fail, a bump 

test fail, 

or 

during an alarm event. To define how often 

lntelliFlash occurs (1-60 seconds), refer to Confidence 

Beep and lntelliFlash Interval. Default setting is 

1 second. Detector is shipped with lntelliFlash enabled. 

Confidence Beep and lntelllFlash Interval: Enter a 

value (1-60 seconds) to define how often lntelliFlash 

occurs and the detector beeps.  lntelliflash and/or 

Confidence Beep must be enabled in order to define 

Confidence Beep and lntelliFlash Interval. 

Language: Select the language to display on the LCD· 

Engli�h. Fran�ls (French), Deutsch (German), Espaflol 

(Spanish), or Portugues (Portuguese). 

GasAlertMicroC/ip 

Charging the Rechargeab, .. Battery 

Charging the Rechargeable Battery 

A Warning 

Only the manufacturer can replace 

the 

rechargeable 

battery. Failure to adhere to this caution can lead to fire 

and/or explosion. 
Charge only in a safe area that is free of hazardous gas 

and within temperatures of 32

°

F to 113

°

F 1o•c to 45°C). 

The charging adapter is specific to your region. Use of the 

charging adapter outside your region will damage the 

charger and the detector. 

Do not calibrate during or Immediately after charging. 

1.  Deactivate the detector. Insert the charging adapter 

plug into an AC outlet. 

Reviews: