7 | Maintenance and Service
16 / 24
Instruction Manual VACTEST DCC 400_EN_en
Settings
Meaning
0
User control
►
sensor has to be switched on and off manually
1 (Default)
Automatic control
►
Automatic sensor control
Manual switch-on and switch-off of the cold cathode can be done by software com-
mand “Cathode Control” (CC) or by applying a defined voltage signal to pin 9 of the
transmitter’s electrical connector. As voltage level 3 … 24 VDC (default) or alternatively
GND can be used, the selection is done by the parameter “Cathode Control Logic” (DL).
Cathode Control Logic (DL):
The cold cathode sensor can be switched on or off by applying a defined voltage signal
to pin 9 of the transmitter’s electrical connector (see chapter
).
By means of the parameter “Cathode Control Logic” (DL) you can select via software
command, if a 24 VDC or GND level signal shall be used.
Settings
Meaning
0
GND
1 (Default)
3 … 24 VDC
The cold cathode sensor will be active as long as the voltage signal is applied.
The parameter “Cathode Control Logic” (DL) can also be queried.
6.3 VacTest Explorer Software
VacTest explorer software has been especially developed for use with Busch vacuum
gauges and is available for Windows and Android operating systems.
VacTest explorer features plotting and saving of measurement data as well as a comfort-
able way of configuring all device parameters.
Download VacTest explorer software on the Busch website
Features example:
– Plot, analyze and save measurement curves.
– Compare multiple plots.
– Export measurement data for MS Excel.
– Automatic calculation of leak rates by rate-of-rise measurements.
– Easy configuration of all device parameters.
– Scaling wizard with graphic support for adjusting the voltage output characteris-
tic.
– The voltage output curve can be modified through the VacTest explorer software
and can directly replace gauges of other brands.
7
Maintenance and Service
WARNING
Units contaminated with hazardous material.
Risk of poisoning!
Risk of infection!
If the unit is contaminated with hazardous material:
• Wear appropriate personal protective equipment.